Most of our pH sensors have a noble metal solution ground and our transmitters measures two values: 1. the voltage between this solution ground, or liquid earth sensor and the pH sensor, and 2. the voltage between this solution ground, or liquid earth sensor and the reference cell. This allows the transmitter to use this voltage to calculate the ORP value. The beauty of this solution is that with one combination pH/ORP sensor and one Transmitter you can get both pH and ORP values.
Related Products & Solutions
-
Liquid Analyzers
- Liquid analyzers
- Monitoring process chemistry
- Water quality
- Process optimization and control
-
pH and ORP Sensors
- pH and ORP sensors
- Proper pH electrode/sensor selection critical for optimal measurement results