In a dual input FLXA202 or FLXA21 analyzer, how can I see the values from the second input module?

In a dual input FLXA202 or FLXA21 analyzer you can use a HART Splitter to convert the HART® dynamic variables into current outputs or contact outputs. The model number is "1W-KFD2-HLC-EX1.D" and the description is "P+F HART Splitter." In the FLXA menu you can select which dynamic variables you like as "SV" "TV" and "FV."

Go to Commissioning>Advanced Setup>Communication>HART> then use the down arrow to select the SV, TV, and FV fields.

Related Products & Solutions

  • 2-Wire Transmitter/Analyzer FLXA21

    The FLEXA™ series analyzers are used for continuous on-line measurements in industrial installations. With an option for single or dual sensor measurement, they are the most flexible two-wire analyzer available.

    See More
  • pH and ORP Analyzers

    pH and ORP meters, analyzers and transmitters are used for continuous process monitoring of pH and ORP to ensure water/product quality, monitor effluent discharge, batch neutralization, pulp stock, scrubbers, cooling towers, chemical, water/wastewater treatment and many other applications.

    See More