Video FAQ
Questions
- Why am I getting a temperature compensation 1 error the analyzer when I'm using either an inductive or contacting conductivity sensor?
- What are the common causes of temperature compensation errors?
Answer
- Make sure the sensor is completely submerged in the process fluid,
- If Step 1 does not clear the code, then most likely you need to clean the sensor. To clean just soak the sensor in a 5 to 10% HCl solution for 5-10 minutes.
Related Products & Solutions
-
Multi Channel 4-Wire Analyzer FLXA402
The FLEXA™ series analyzers are modular-designed analyzers used for continuous online measurements in industrial installations. They offer single or multi-sensor measurement.
-
Conductivity Sensors
- Conductivity sensors and electrodes
- Process conductivity, resistivity, WIFI, demineralizer water, RO water, percent concentration, boiler blowdown, and TDS
- Retractable, flow thru, immersion, and direct insertion
-
Liquid Analyzers
- Liquid analyzers
- Monitoring process chemistry
- Water quality
- Process optimization and control