| Measurement object |
Free chlorine or residual chlorine (total chlorine) in water |
| Measurement method |
Rotating electrode type polarographic system |
| Measurement range |
0 to 10 mg/l (mg/l used as concentration units per JIS K0101 and municipal water test methods) |
| Output range |
Can be freely set within ranges with spans of 1 mg/l or greater |
| Output signal |
4 to 20 mA DC (load resistance, 550 Ω Max.) or 1 to 5 V DC (output resistance, 300 Ω Max.) |
| Sample conditions |
Temperature: 0 to 50 deg C
pH: 3 to 9
Flow: 1 to 4 l /min, for pure water or municipal water application.
5 to 10 l /min, for raw water or sea water application.
10 to 20 l /min, for secondary sewage treatment application.
Pressure: 20 to 500 kPa |
| Electrodes |
Measuring electrode: Rotating platinum electrode
Reference electrode: Platinum electrode (built-in Pt1000 Ω RTD) |
| Ambient Temperature |
-5 to 55 deg C |
| Ambient Humidity |
5 to 95 % RH (non-condensing) |
| Installation |
Indoors (A separate rainproof cover is required for outdoor installation. Avoid direct sunlight.) |
| Power Supply |
100/110/220 V AC ±10 %, 50/60 Hz |
| Weight |
Approx. 65 kg (for pure water) |