| Measurement |
| Either ammonia gas or acidic gas (anion analysis: hydrogen chloride, suifur dioxide, etc.) |
|
| Gas collection method |
Diffusion scrubber system |
| Analysis method |
Ion chromatography |
| Measuring range |
0 to 50 μg/m3 |
| Minimum detection limit |
0.01 μg/m3 (Ammonia), 0.02mg/m3 ( Acid gas) |
| Measurement cycle |
Minimum 35 min. cycle. |
| Measuring points |
One point or up 16 points (optional) |
| Calibration |
Calibration using standard solutions, automatic calibration |
| External contact input |
Measuring stream skip signal, external alarm signal, external alarm reset signal |
| External contact output |
Abnormal concentration signal, system alarm signal, measuring stream signal, re-measurement signal, liquid leakage signal, alarm buzzer signal, alarm light signal |
| Power supply |
100 to 120 V AC, 50/60 Hz or 200 to 240 V AC, 50/60 Hz |
| Weight |
Approx. 170kg (except for chemical solution, workstation, and pure water system.) |
|