| Measurement |
| Ammonia; NH3 |
| Cations; Na+, NH3, K+, Ca2+, Mg2+ |
| Anions; Cl-, Br-, NO2-, NO3-, SO32-, SO42-, PO43- |
| Fluoride/organic acids ;F-, HCOO-, CH3COO-, Cl- |
| Boron; B |
|
| Gas collection method |
Impinger method |
| Analysis method |
Ion chromatography |
| Measuring range |
| NH3, Cations, Anions; 0 to 50 μg/m3 |
| Fluoride/organic acid; F-, Cl-: 0 to 5 μg/m3, HCOO-, CH3COO-: 0 to 30 μg/m3 |
| B; 0 to 100ng/m3 |
|
| Minimum detection limit |
| NH3, Cations: 0.01 μg/m3, Acid gas; 0.02 μg/m3, PO4M3-: 0.05 μg/m3, F-, Cl-: 0.01 μg/m3 HCOO-: 0.05 μg/m3, CH3COO-: 0.01 μg/m3, B: 10ng/m3 |
|
| Measurement cycle |
Mini60 min. or 90 min. |
| Measuring points |
One point or up to 16 points (optional) |
| Calibration |
Calibration using standard solutions, automatic calibration |
| External contact input |
Measuring stream skip signal, external alarm signal, external alarm reset signal |
| External contact output |
Abnormal concentration signal (for every stream, relay contact),
system alarm signal (relay contact), measuring stream signal (binary, relay contact), re-measurement signal (relay contact), liquid leakage signal (relay contact), alarm buzzer signal, alarm light signal |
| Power supply |
100 to 120 V AC, 50/60 Hz |
| Weight |
Approx. 150to 180kg depen on specifications |
|