- Home>
- Products & Solutions:
- Environmental & Analytical Products:
- Conductivity Analyzers:
- SC450G 4-Wire Conductivity/ Resistivity Analyzer:
- DC402G Dual Cell Conductivity Analyzer:
- FLXA21 Modular 2-Wire Conductivity Analyzer:
- SC202G 2-Wire Conductivity/ Resistivity Analyzer:
- SC72 Personal Conductivity Meter:
- ISC450G 4-Wire Inductive Conductivity Analyzer:
- FLXA21 Modular 2-Wire Inductive Conductivity Analyzer:
- ISC202G 2-Wire Inductive Conductivity Analyzer:
- SC210G Conductivity Sensor:
- SC8SG Conductivity Sensor:
- SC4AJ Conductivity Sensor:
- SC42, FF40/FS40/FD40:
- ISC40GJ, ISC40S Inductive Conductivity Sensors>
- FLXA21 Modular 2-Wire Inductive Conductivity Analyzer
FLXA21 Modular 2-Wire Inductive Conductivity Analyzer
Product Information

The FLXA21 two-wire liquid analyzer offers single or dual sensor measurement. The modular-designed analyzer offers 4 kinds of measurements – pH/ORP (oxidation-reduction potential), contacting conductivity (SC), inductive conductivity (ISC) or dissolved oxygen (DO) – with the respective sensor module.
The FLXA21 Inductive Conductivity Analyzer performs non-contact measurement according to the inductive measurement principle so that it can be used to monitor the concentrations of acid, alkali and saline solutions in the chemical industry, as well as to measure the conductivity in a wide range of applications in industries such as food, pharmaceutical industry, plating and surface treatment processes, and paper and pulp. In addition, it has matrix compensation and output linearization functions to accurately measure strong acid and alkali solutions. Only one sensor can be connected.
For the wide range of industrial environment, the FLXA21 Inductive Conductivity Analyzer is designed with the enclosure of plastic, stainless steel or stainless steel with corrosion resistant coating.
| Bulletin for FLXA21 |
| General Specification for FLXA21 |
| General Specification for Inductive Conductivity Sensors |
| User's Manual for FLXA21 |
| Selection Guide for Conductivity Analyzer |
Features
| * | Easy touch screen operation on 2-wire type analyzer |
| * | Simple HMI menu structure in 12 languages (English, Chinese, Czech, French, German, Italian, Korean, Polish, Portuguese, Russian, Spanish and Japanese) |
| * | Quick setup menu for immediate measurement |
| * | Indication of sensor wellness |
| * | HART communications |
| * | Enclosure – plastic, stainless steel or stainless steel with corrosion-resistant coating |
Specifications
| Input specifications | Compatible with ISC40 inductive conductivity sensor with integrated temperature sensor: 30kΩ or Pt1000. |
| Input range | Conductivity: 0 to 1999 mS/cm at 25℃ (77℉) Minimum: 1 μS/cm (at process temperature) Maximum: 2 S/cm (at process temperature) Temperature: -20 to 140℃ (0 to 280℉). |
| Cable length | Max. 60 meters (200 feet). 10 meters (35 feet) fixed sensor cable + 50 meters (165 feet) WF10 extension cable. |
| Transmission signal | One output of 4 to 20 mA DC. Bi-directional HART® digital communication, superimposed on mA1 (4 to 20mA) signal. Burn up (22.0mA) or burn down (3.9mA) to signal failure accorded to NAMUR NE43. |
| Temperature compensation | Automatic or manual, for temperature ranges. Reference temp. programmable from 0 to 100℃ or 30 to 210℉ (default 25℃). |
| Calibration | Semi-automatic calibration using pre-configured OIML (KCl) stanard tables, with automatic stability check. Manual adjustment to grab sample. |
| Display | Black/White (213 x 160 pixels) LCD with a touch screen. |
| Analyzer Structure | Module structure |
| Housing |
Case:
Plastic (Polycarbonate), stainless steel without painting, stainless
steel with epoxy coating, stainless steel with urethane coating
Window: Polycarbonate (flexible) Protection: NEMA4X, IP66 |
| Environment | Ambient Temperature: -20 to 55℃ Storage Temperature:-30 to +70℃ Humidity: 10 to 95 % RH (Non-condensing) |
| Regulatory Compliance |
Safety :
EN61010-1, CSA C22.2 No.61010-1 EMC : EN61326-1 Class A,Table 2 (For use in industrial locations), EN61326-2-3, AS/NZS CISPR11, Korea Electromagnetic Conformity Standard Installation altitude: 2000 m or less Category based on IEC 61010: I Pollution degree based on IEC 61010: 2 |
| Explosion-proof | ATEX, IECEx, FM, CSA |
| Power Supply | Nominal 24 V DC loop powered system |
| Conduit Adapter | G1/2, 1/2NPT, M20 x 1.5 |
| Weight | Approx. 1 kg (Plastic housing), Approx. 2 kg (Stainless steel housing) |
Performance (Accuracy)
(The specifications are expressed with simulated inputs.)
(Output span is 0-100 μS/cm or more)
| Conductivity |
Linearity: ±(0.4 %F.S. + 0.3 μS/cm)
Repeatability: ±(0.4 %F.S. + 0.3 μS/cm) |
| Temperature | ±0.3℃ |
| Step response | 90 % (< 2 decades) in 8 seconds |