Category Sitemap
Home / Our Businesses / IA / Products / Environmental.. / Conductivity .. /
Conductivity Meter SC402G
Converter for Conductivity or Resistivity
Model SC402G

Flexibility, reliability and low maintenance are among the benefits provided by the EXA SC402G conductivity analyzer. Designed to meet the exacting requirements of measuring conductivity and resistivity in the modern industrial environment, it contains many features to ensure the best precision whatever the application.

This 4-wire converter is housed in a robust IP65 field mountable case. Two mA outputs, four relays and a clear LCD make the SC402G a truly comprehensive package.

The SC402G features PI control on both the auxilliary mA output and the pulse proportional relay outputs, thus avoiding the need for a separate controller.

A wide variety of temperature compensation possibilities (NaCl according to IEC 746-3, manual TC, preprogrammed matrices and a free programmable 5x5 matrix) provide a high-accuracy measurement with minimum effort.

* Universal conductivity/resistivity, software switchable
* On-line sensor checking
* Matrix temperature compensation for pure water applications
* Four fully configurable SPDT contact outputs
* Two fully configurable mA outputs
* Built-in PI controller
* Easy to use EXA control panel
* Pharmaceutical water monitoring (USP23)

Input specifications Two or four electrode measurement with cell constant from 0.008 to 50.0 cm -1
Detection method Frequency, read-pulse position and reference voltage are dynamically optimized.
Input ranges
- Conductivity : 0.000 micro S/cm to 1999 mS/cm at 25 degrees C (77 degrees F) reference temperature.
  Minimum : 1 micro S x C at process temperature
(underrange 0.000 micro S/cm).
Maximum : 500 mS x C at process temperature
(overrange 1999 mS x C).
- Resistivity : 0.000 k ohm - 999 M ohm/C at 25 degrees C (77 degrees) reference temperature.
  Minimum : 0.002 k ohm/C at process temperature
(underrange 0.000 k ohm x cm).
Maximum : 1 M ohm /C at process temperature
(overrange 999 M ohm x cm).
- Temperature
Pt1000 : -20 to +250 degrees C (0 - 500 degrees F)
Pt100 and Ni100 : -20 to +200 degrees C (0 - 400 degrees F)
8K55 NTC : -10 to +120 degrees C (10 - 250 degrees F)
PB36 NTC : -20 to +120 degrees C (0 - 250 degrees F)
Span
- Conductivity : - min. 0.010 micro S/cm,
- max. 1999 mS/cm. (max 90% zero suppression)
- Resistivity : - min. 0.001 k ohm x cm
- max. 999 M ohm x cm. (max 90% zero ppression)
- Temperature : Dependent on temp. sensor
Sensor type min. max.
Pt1000 25 degrees C
(50 degrees F)
250 degrees C
(500 degrees F)
Pt100, Ni100 25 degrees C
(50 degrees F)
200 degrees C
(400 degrees F)
PB36 NTC, 8K55 NTC 25 degrees C
(50 degrees F)
100 degrees C
(200 degrees F)
The instrument is user programmable for linear or non-linear conductivity ranges.
Transmission Signals Two isolated outputs of 0/4-20 mA
DC with common negative.
Maximum load 600 ohm.
Auxiliary output can be chosen from conductivity, resistivity,
temperature or PI control Burn up (22 mA) or Burn down (0/ 3.5 mA) to signal failure.
Temperature compensation
Automatic, for temperature ranges mentioned under C (inputs).
- Reference temp. : programmable from 0 to 100 degrees C or
30 - 210 degrees F (default 25 degrees C).
Compensation algorithm
According IEC 746-3 NaCl tables (default).
Two independent user programmable temperature coefficients,
from 0% to 3.5% per degree C (degree F) by adjustment or calibration.
- Matrix compensation : with conductivity function of concentration and temperature.
Choice of 5 preprogrammed matrixes and a 25-point user programmable matrix.
Display Custom liquid crystal display, with a main display of 3 1/2 digits 12.5
mm high. Message display of 6 alphanumeric characters, 7 mm
high. Warning flags and units (mS/cm, k ohm.cm, micro S/cm and
M ohm.cm) as appropriate.
Contact Outputs
- General : Four (4) SPDT relay contacts with LED indicators. For S1, S2, and S3, the LED is on when relay is powered.
Contact outputs configurable for hysteresis and delay time.
- Switch capacity : Maximum values 100 VA, 250 VAC, 5 Amps.
Maximum values 50 Watts, 250 VDC, 5 Amps.
- Status : High/low process alarms, selected from conductivity, resistivity and temperature.
Contact output is also available to signal "Hold active"
- Control function : On / Off
PI pulsed - Proportional duty cycle control with integral term.
PI frequency - Proportional frequency control with integral term.
In addition FAIL alarm for system and diagnostic errors on S4
Contact Input
Remote range switching to 10 times the programmed range.
- Contact open : If impedance > 100 k ohm: Range 1
- Contact closed : If impedance < 10 ohm: Range 2 (10 x Range 1)
Power Supply - 230 VAC +/- 15%, 50/60 Hz, maximum consumption 15 VA.
- 115 VAC +/- 15%, 50/60 Hz, maximum consumption 15 VA.
- 100 VAC +/- 15%, 50/60 Hz, maximum consumption 15 VA.
Input isolation 1000 VAC or 2100DC
Shipping Details Package size w x h x d
290 x 225 x 170 mm.
11.5 x 8.9 x 6.7 in.
Packed weight approx. 2.5 kg (5lb).

GLOBAL  
YOKOGAWA

Powered by Ultraseek
·  Conductivity Meter SC402G