| Measuring object |
Gas density of all gases except for corrosive and acetylene gas |
| Measuring principle |
Vibration type (resonant frequency) |
| Sample gas conditions |
| Temperature: -10 to 50 deg C (non-condensing) |
| Pressure: Max. 588.4 kPa (abs) |
| Gas flow: 0.1 to 1 l/min |
|
| Measuring range |
Gas density: 0 to 6 kg/m3 |
| Output signal |
| Output 1: 4-20 mA DC, Isolated, load resistance: 600Ω max. |
| Output 2: 4-20 mA DC, Isolated, load resistance: 600Ω max. |
|
| Explosionproof approval Converter |
Rainproof for outdoor use, equivalent to IP65 / NEMA 4X
Ambient temperature :-10 to 55 deg C
Ambient humidity : 5 to 95%RH
| GD402T: |
FM Explosionproof Approval.
Explosionproof for Class I, Division 1, Groups B, C and D;
Dust Ignitionproof for Class II, III, Division 1, Groups E, F and G.
Enclosure : NEMA Type 4X
Temperature code : T6
Electrical connection: 1/2NPT female |
| GD402V: |
CSA Explosionproof Approval.
Explosionproof for Class I, Division 1, Groups B, C and D;
Dust Ignitionproof for Class II, III, Division 1, Groups E, F and G.
Enclosure : Type 4X
Temperature code : T6
Electrical connection: 1/2NPT female |
| GD402R:: |
TIIS Explosioproof approval.
Explosionproof code : Exd IIB+H2T6
Temperature code : T6
Electrical connection: G3/4 female |
|
| Explosionproof approval detector |
Rainproof for outdoor use, equivalent to IP65 / NEMA 4X
Ambient temperature :-10 to 60 deg C
Ambient humidity : 5 to 95%RH
| GD40T: |
FM Explosionproof and Intrinsically safe Approval.
Explosionproof for Class I, Division 1, Groups B, C and D;
Dust Ignitionproof for Class II, III, Division 1, Groups E, F and G with Intrinsically Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
Enclosure : NEMA Type 4X
Temperature code: T5
Electrical connection : 1/2NPT female
Process connection : 1/4NPT female |
| GD40V: |
CSA Explosionproof and Intrinsically safe Approval.
Explosionproof for Class I, Division 1, Groups B, C and D;
Dust Ignitionproof for Class II, III, Division 1, Groups E, F and G with Intrinsically Safe sensor for Class I, II, III, Division 1, Groups
B, C, D, E, F and G.
Enclosure : Type 4X
Temperature code : T5
Electrical connection : 1/2NPT female
Process connection : 1/4NPT female |
| GD40R: |
TIIS Explosionproof and Intrinsically safe Approval.
Explosionproof code : Exd [ia] IIB+H2T5
Temperature code : T5
Electrical connection : G1/2 female
Process connection : Rc1/4 female
|
|
| Power supply |
100 to 240 V AC, 24V DC, 50/60 Hz |
|