| Measured gas |
O2 in N2 or mixed gases; cannot be used with silicone vapor, combustible gases and organic solvent |
| Measurement principle |
Zirconia current limiting method |
| Measuring range |
0 to 25% |
| Sampling method |
Natural diffusion system. If the sampled gas may include organic solvents, use a sampling unit. |
| Measuring gas conditions |
Atmospheric pressure, operating temperature; 0 to 70 degC, gas flow up to 1 m/s |
| Output signal : 2 outputst |
4-20 mA DC automatic/manual switching mode, 1 to 5 V DC fixed (0-100 ppm/0-1000 ppm/0-10000 ppm/0-25%) |
| Output range |
0-100 ppm/0-1000 ppm/0-10000 ppm/0-25%, automatic/fixed switching mode |
| Contact output |
2 contact outputs |
| Calibration |
a. Atmospheric air one-point calibration : 15 to 25% O2 range
b. One-point calibration at 100 ppm : 80 to 120 ppm O2 range
c. One-point calibration at 1000 ppm : 900 to 1100 ppm O2 range
d. One-point calibration at 10000 ppm : 8500 to 9500 ppm O2 range
e. Complete calibration : with air, 20, 100, 500, 1000, 10000 ppm in that order |
| Extension cable |
4 m, 9 m, 29 m |
| Power supply |
100 to 240 V AC, 50/60 Hz, approx. 15 W |
|