Explosionproof Zirconia Oxygen Analyzers
Models ZR22S, ZR202S


Explosionproof Zirconia Oxygen Analyzers
Regulatory Compliance

Two types are available for an explosionproof direct in situ zirconia oxygen analyzer. Model ZR22S/ZR402G is a separate type which consists of a ZR22S explosionproof probe and a ZR402G non-explosionproof converter.

Model ZR202S is an integrated type which combines a probe and a converter. Both Zirconia oxygen analyzers do not need a sampling device, and allow direct installation of the probe in the wall of a flue or furnace to measure the concentration of oxygen in the stack gas. The converter displays the cell temperature and cell emf in addition to the oxygen concentration.

This analyzer is most suitable for monitoring combustion and controlling the low-oxygen combustion of various industrial furnaces in such explosive atmospheres as at petroleum refineries, petrochemical plants, and natural gas plants.


   Bulletin for ZR22S, ZR202S 

   General Specification for ZR22S, ZR202S 

   User's Manual for ZR22S/ZR402G 
   User's Manual for ZR202S 
* The built-in heater assembly of the probe can be replaced on site, reducing maintenance costs.
* The probe uses a long-life, high-reliability Zirconia sensor.
* The separate type converter ZR402G incorporates a LCD touch-screen for ease of operation.
* The integrated type ZR202S integrates both probe and converter, to reduce wiring, piping, and installation costs. ZR202S of unit uses an optical switch for ease of operation at the site.
* Remote maintenance using digital communications (HART) reduces maintenance costs.
Note: HART is a registered trademark of HART Communication Foundation.
ZR22S/ZR402G
Measurement object Oxygen concentration in combustion exhaust gas and mixed gas.
Measurement system Zirconia system
Explosionproof Approval ZR22S-A (ATEX); EExd II B + H2, Group II,Category 2GD, T2, T300 deg C
ZR22S-B (FM); Class I, Division1, Groups B, C and D, Class II/III, Division1, Groups E, F and G, T2
ZR22S-C (CSA); Class I, Division1, Groups B, C and D, Class II/III, Division1, Groups E, F and G, T2
ZR22S-D (IECEx); Exd II B + H2, T2, ExtD A21 IP66 T300 deg C
Oxygen concentration 0.01 to 100 vol% O2
Output signal 4 to 20 mA DC, two points (maximum load resistance 550Ω)
Measurement range Any setting in the range of 0 - 5 vol%O2 to 0 - 100 vol%O2 (in 1 vol% O2), or partial range.
Digital communication (HART) 250 to 550Ω, depending on number of field devices connected to the loop (multi-drop mode)
Sample gas pressure -5 to +5 kPa
For 0.15m probe, -0.5 to +5 kPa.
No pressure fluctuation in the furnace should be allowed.
Sample gas temperature 0 to 700 deg C(Probe only)
700 to 1400 deg C (with 0.15 m probe and high temperature probe adapter)
Probe length 0.15, 0.4, 0.7, 1.0, 1.5, 2.0 m
Contact output (ZR402G) Four points (one is fail-safe, normally open)
Contact input (ZR402G) Two points
Calibration (ZR402G) Method; zero/span calibration
Calibration mode; automatic, semi-automatic and manual (with an LCD touch screen)
Construction of detector Equivalent to NEMA4X/IP66
Construction of converter Outdoor installation, equivalent to NEMA4X/IP66
Ambient Temperature (ZR22S) -20 to +60 deg C (-20 to +150 deg C on the terminal box surface)
Ambient Temperature (ZR402G) -20 to +55 deg C
Reference Air System Instrument Air; 50 kPa + the pressure inside the furnace Consumption; Approx. 1Nl/min
Wiring Connection ATEX; M20 by 1.5 mm or 1/2 NPT select one type (2 pieces)
FM; 1/2 NPT (2 pieces)
CSA; 1/2 NPT (2 pieces)
IECEx; M20 by 1.5 mm or 1/2 NPT select one type (2 pieces)
Power supply 85 to 264 V AC; 45 to 66 Hz


ZR202S
Measurement object Oxygen concentration in combustion exhaust gas and mixed gas.
Measurement system Zirconia system
Display 6-digit LCD
Switch Three optical switches
Explosionproof Approval ZR202S-A (ATEX); EExd II B + H2, Group II, Category 2GD, T2, T300 deg C
ZR202S-B (FM); Class I, Division1, Groups B, C and D, Class II/III, Division1, Groups E, F and G, T2
ZR202S-C (CSA); Class I, Division1, Groups B, C and D, Class II/III, Division1, Groups E, F and G, T2
ZR202S-D (IECEx); Exd II B + H2, T2, ExtD A21 IP66 T300 deg C
Oxygen concentration 0.01 to 100 vol% O2
Output signal 4 to 20 mA DC, one point (maximum load resistance 550Ω)
Measurement range Any setting in the range of 0 - 5 vol%O2 to 0 - 100 vol%O2 (in 1 vol% O2), or partial range.
Digital communication (HART) 250 to 550Ω, depending on number of field devices connected to the loop (multi-drop mode)
Sample gas pressure - 5 to + 5 kPa
No pressure fluctuation in the furnace should be allowed.
Sample gas temperature 0 to 700 deg C
Probe length 0.4, 0.7, 1.0, 1.5, 2.0 m
Contact output Two points (one is fail-safe, normally open)
Contact input Two points
Calibration Method; zero/span calibration
Calibration mode; automatic, semi-automatic and manual (with optical switch)
Construction Equivalent to NEMA4X/IP66
Ambient temperature -20 to +55 deg C (- 5 to +70 deg C on the case surface)
Power Supply 85 to 264 V AC, 45 to 66 Hz
Repeatability ±0.5% Maximum value of set range
Linearity ±1% Maximum value of set range
Drift ±2% Maximum value of set range/month
Response time 90% response within 5 sec