Category Sitemap
Home / Our Businesses / IA / Products / Environmental & Analytical Products / pH/ORP Analyzers /
pH/ORP Sensors PH8EFP, PH8ERP, OR8EFG, OR8ERG
pH/ORP Sensors
Models PH8EFP, PH8ERP, OR8EFG, OR8ERG


pH/ORP Sensors

Yokogawa's pH and ORP analyzers are highly reliable. They feature advanced functions and are useful for a wide variety of applications including water quality management in a broad range of production processes and in medium-sized wastewater treatment plants, or as general pH and ORP control systems.

Based on Yokogawa's track record and years of experience, a comprehensive range of products has been produced to provide solutions best suited to individual applications.

In addition to the PH8EFP, PH8ERP, OR8EFG and OR8ERG standard Ryton pH and ORP sensors, specialty pH and ORP sensors are available for various applications.


   General Specification  

   User's Manual for PH8EFP  
   User's Manual for PH8ERP  
   User's Manual for OR8EFG  
   User's Manual for OR8ERG  
Ryton pH/ORP sensors PH8EFP, PH8ERP, OR8EFG, OR8ERG

* With a body made of Ryton, a strong engineering plastic which is comparable to Teflon in terms of corrosion resistance and heat resistance, a wide range of applications are possible.
* A single type of electrode can support all applications regardless of whether a holder or cleaner is used.
* The integrated-electrode design simplifies calibration with standard solutions and maintenance.
* The glass electrode of a pH analyzer, the platinum or gold electrode, and the junction of an ORP analyzer can each be easily replaced.
1. pH Sensors
1-1. KCl Refillable Type Electrode PH8ERP
Measured object Hydrogen ion concentration (pH) in aqueous solution
Measurement principle Glass electrode method
Measuring range pH 2 to 12
Measuring temperature -5 to 80 deg C
Measuring pressure Atmospheric pressure to 50 kPa
Temperature compensation sensor Pt1000
Weight Approx. 0.4kg

1-2. KCl Filling Type Electrode PH8EFP
Measured object Hydrogen ion concentration (pH) in aqueous solution
Measurement principle Glass electrode method
Measuring range pH 0 to 14
Measuring temperature -5 to 105 deg C
Measuring pressure Atmospheric pressure to 10 kPa (General purpose or big volume tank 500 ml)
Atmospheric pressure to 500 kPa (Medium pressure)
Temperature compensation sensor Pt1000
Weight Approx. 0.4 kg

2. ORP Sensors
2-1. KCl Refillable Type Electrode OR8ERG
Measured object Oxidation-Reduction potential in aqueous solution
Measurement principle Metal electrode method
Measuring range -1500 to 1500 mV
Measuring temperature -5 to 80 deg C
Measuring pressure Atmospheric pressure to 50 kPa
Weight Approx. 0.4 kg

2-2. KCl Filling Type Electrode OR8EFG
Measured object Oxidation-Reduction potential in aqueous solution
Measurement principle Metal electrode method
Measuring range -1500 to 1500 mV
Measuring temperature -5 to 105 deg C
Measuring pressure Atmospheric pressure to 10 kPa (General purpose or big volume tank 500 ml)
Atmospheric pressure to 500 kPa (Medium pressure)
Weight Approx. 0.4kg