| Measurement |
Up to 4 components (NOx, SO2, CO2, CO) and O2 |
| Measurering method |
| NOx, SO2, CO2, CO: Non-dispersive Infrared Absorption (NDIR) |
| O2: Zirconia or paramagnetic Cell method |
|
| Measuring ranges |
| NOx: |
:0-50 ppm to 0-5000 ppm |
| SO2: |
0-50 ppm to 0-1000 ppm |
| CO2: |
0-1 vol% to 0-20 vol% |
| CO: |
0-50 ppm to 0-5000 ppm |
| O2: |
0-10/0-25 vol% |
|
| Display |
LCD with backlight |
| Output signal |
5 points, 4-20mA DC or 0 to 1 V DC |
| Contact output |
Each component range identification, analyzer failure, calibration failure, calibration status, maintenance status, CO peak count alarm, each instantaneous value alarm, analyzer power shutdown |
| Contact input |
Auto calibration start, average reset, Range switching output hold, remote pump off |
| Construction |
Outdoor/indoor stand-alone system, non-explosionproof, rainproof |
| Ambient temperature |
-5 to 40 degC |
| Ambient humidity |
90%RH or less |
| Power supply |
100/110/115/200/230 V AC ±10%, 50/60 Hz ±0.5 Hz |
| Power consumption |
Approx. 600 to 1000 VA |
| Weight |
Approx. 350 kg (without standard gases) |
| Sample gas conditions |
| Temperature: 1400 degC or less |
| Dust: 500 mg/Nm3 or less |
| Pressure: -5 to 5 kPa |
| Flow rate: Approx. 2 l/min |
|
|