Rotameter and Trogamid or Polusulfon
Product Information

Assembling Position: Vertical
Assembly: Before installing the device in the tube remove all packaging particles and the cardboard strip which serves to secure the float during transport. The existing tube must be exactly in line, so that the rotameter can be installed tension-free.
Fasten bolted joints only until the rotameter is tight. Usually fastening by hand is enough; maximum torque must not exceed approx. 10 Nm.
Leakage check: Do not use any leak test fluids (sprays) containing alcohol or solvents. We recommend soap and water solution to check for leaks.
Commissioning: Please make sure that there is no alcohol or solvent in the measuring fluid.
Maximum pressure load: 20 bar abs. with iron couplings 10 bar abs. with plastic couplings
Maximum temperature: - 10° up to +100° C
Maintenance: The RAQN rotameters are maintenance-free. If dirt accumulates in the meter tube, remove the device and clean it with soap and water solution (not with solvent or alcohol). Remove dirt or dust on the outside with a dustcloth.
Assembly: Before installing the device in the tube remove all packaging particles and the cardboard strip which serves to secure the float during transport. The existing tube must be exactly in line, so that the rotameter can be installed tension-free.
Fasten bolted joints only until the rotameter is tight. Usually fastening by hand is enough; maximum torque must not exceed approx. 10 Nm.
Leakage check: Do not use any leak test fluids (sprays) containing alcohol or solvents. We recommend soap and water solution to check for leaks.
Commissioning: Please make sure that there is no alcohol or solvent in the measuring fluid.
Maximum pressure load: 20 bar abs. with iron couplings 10 bar abs. with plastic couplings
Maximum temperature: - 10° up to +100° C
Maintenance: The RAQN rotameters are maintenance-free. If dirt accumulates in the meter tube, remove the device and clean it with soap and water solution (not with solvent or alcohol). Remove dirt or dust on the outside with a dustcloth.

The powerful line-up of DPharp digital family is launched.
The world's smallest dual bent tube Coriolis Mass Flowmeter


The highlights of DPharp unique technologies that provides superior performance, stability and reliability in your measurement.
- EJX B Series Differential Pressure/Pressure Transmitters
Yokogawa's versatile field device management tool FieldMate is a so called hybrid solution supporting both EDDL as well as FDT/DTM technology.