- Home>
- Products & Solutions:
- Field Instruments:
- Vortex Flowmeters:
- digitalYEWFLO Vortex Flowmeters:
- digitalYEWFLO Reduced Bore Type Vortex Flowmeter:
- digitalYEWFLO Multivariable Type Vortex Flowmeter:
- digitalYEWFLO Vortex Flowmeter High process Temperature Version / Cryogenic Version>
- digitalYEWFLO Fieldbus Type Vortex Flowmeter:
- digitalYEWFLO Vortex Flowmeters
High Process Temperature Version / Cryogenic Version
digitalYEWFLO Vortex Flowmeters
For high temperature or cryogenic flow measurement.
Product Information

![]() |
||||||||||||||||||||||||||
| We have the process temperature version and cryogenic version in the digitalYEWFLO series which boast the long-term reliability. The high process temperature version is available up to 450 degrees C, the highest temperature in the vortex flowmeter. The cryogenic version is available up to -196 degrees C . These versions are the best solutions for the high temperature steam application or the cryogenic liquid oxygen up to 200 mm of the nominal sizes. |
||||||||||||||||||||||||||
|
||||||||||||||||||||||||||
Features: High Process Temperature Version / Cryogenic Version
| - | New functions with Spectral Signal Processing (SSP) technology |
| SSP analyses the fluid conditions inside the digitalYEWFLO vortex flowmeter and uses the data to automatically select the optimum adjustment for the application, providing features never seen before in a vortex flowmeter. SSP accurately senses vortices in the low flow range, providing outstanding flow stability. |
|
| - | Advanced self-diagnostics |
| Application conditions such as high pipeline vibration and abnormal flow are predicted and indicated. | |
| - | High accuracy |
| ±0.75% of reading (liquid) ±1.0% of reading (gas, steam) |
|
| - | Wide process temperature range |
| High temperature version up to 450 deg C Cryogenic version minimum -196 deg C |
|
| - | Simple parameter settings |
| Frequently-used selections grouped together in a quick-access format decrease commissioning time | |
| - | Clear, concise indicator |
| Convenient display of the simultaneous and total flow rates together with the process diagnosis | |
| - | Dual output for analog/pulse |
| Simultaneous output for flow rate and pulse | |
| - | Alarm output |
| An alarm signal is output when an alarm occurs. | |
| - | BRAIN/HART communication and key switches for human-machine interface |
| - | NACE compliant |
| - | NAMUR 43 compliant |
| - | Remote cable length up to 30m |
Specifications:High Process Temp. Version / Cryogenic Version
| Standard | Multi-Variable | Reduced Bore | |
| Model & code | DY Vortex Flowmeter | Optional code:/MV | Option code:/R1, /R2 (Release in the end of September) |
| Fluid to be measured | Liquid, gas, steam (avoid multiphase flow and sticky fluids) | ||
| Nominal size | 15mm ∼ 300mm (400mm:special) |
25mm ∼ 200mm | 25mm ∼ 200mm |
| Accuracy | Liquid: ±1.0% of reading(20000≤Re≤1000) Liquid: ±0.75% of reading(D×1000≤Re) |
Liquid: ±1.0% of reading (20000≤Re≤1000) |
|
| Gas, steam: ±1.0% of reading (flow velocity less than 35m/s) Gas, steam: ±1.5% of reading (flow velocity 35m/s~80m/s) |
|||
| Output signal | Simultaneous output(both analog and transistor contact output available) | ||
| Process temperature range |
-29 deg C ∼ 260 deg C(standard) -196 deg C ∼ 100 deg C(option: cryogenic version) -29 deg C ∼ 450 deg C(option: high process temperature version) |
||
| Process pressure limit |
-0.1MPa (-1kg/cm2) to flange rating | ||
| Ambient temperature | -29 to 85 deg C (Integral type / No indicator) -29 to 80 deg C (Integral type / with indicator) -29 to 85 deg C (Remote type detector) -40 to 85 deg C (Remote type converter / No indicator) -30 to 80 deg C (Remote type converter / with indicator) |
||
| Ambient humidity |
5~100%RH( Non-condensing) | ||
| Mounting | Integral type and Remote type detector: Flange mounting or wafer mounting Remote type converter: 2 inch pipe mounting JIS10/20/40、ANSI150/300/600/900, JPI150/300/600,DIN PN10/16/25/40 |
||
| Electrical connection | JIS G1/2 female, ANSI 1/2NPT female, ISO M20×1.5 female | ||
| Electrical classification | JIS flameproof, FM explosion proof/intrinsically safe ATEX (CENELEC flameproof/intrinsically safe/IEC type n) CSA explosion proof/intrinsically safe : Dual Seal SAA flame proof/intrinsically safe |
||
| Material | Body: SCS14A cast stainless steel (equivalent to SUS316,CF8M) Shedder bar: duplex stainless steel (option:anti-corrosion version) Gasket: SUS316 stainless steel with polytetrafluoroethylene (Teflon) coating Converter housing and case cover: aluminum alloy |
||
| Applicable Standard | *EMC Conformity Standards:
EN61326-1 Class A, Table 2 (For use in industrial locations), EN61326-2-3 EN55011 Class A Group 1 *Pressure Equipment Directive: Notified Body Identification Number 0038 Module H |
||
| Please see Features of digitalYEWFLO for common items. |