EJA430A Gauge Pressure Transmitter |
| A Capsule |
B Capsule |
| Range |
-0.1 to 3 MPa
(-15 to 430 psi) |
-0.1 to 14 MPa
(-15 to 2000 psi) |
| Span |
0.03 to 3 MPa
(4.3 to 430 psi) |
0.14 to 14 MPa
(20 to 2000 psi) |
| Accuracy |
±0.065% |
±0.065% |
| Degrees of protection |
IP67, NEMA 4X, and JIS C0920 immersion proof |
| Certificates |
FM, CENELEC ATEX, CSA, IECEx |
| Output |
4 to 20 mA DC or FOUNDATION™ fieldbus or PROFIBUS PA 2-wire system with digital communication |
| Supply voltage |
BRAIN and HART:10.5 to 42 V DC (10.5 to 30 V DC for Intrinsically safe type)
Fieldbus:9 to 32 V DC (9 to 24 V DC for Entity model and 9 to 17.5 V DC for FISCO model) |
| Ambient temperature |
-40 to 85 deg C (-40 to 185 deg F) (general use type)
-30 to 80 deg C (-22 to 176 deg F) (with integral indicator) |
| Process temperature |
-40 to 120 deg C (-40 to 248 deg F) (general use type) |
| Maximum overpressure |
4.5 MPa (645 psig) |
21 MPa (3000 psig) |
| Mounting |
2-inch pipe mounting |
| Wetted parts material |
Capsule |
SUS316L(Diaphragm material is Hastelloy C-276), Hastelloy C-276, Tantalum, Monel |
| Cover flange |
SUS316, Hastelloy C-276, Monel |
| Housing |
Cast aluminum alloy or SUS316 equivalent stainless steel (optional) |