Question
Common Items
CA11E / CA12E Handy Calibrators
CA71 Handy Calibrator
CA150 Handy Calibrator
Answer
Common Items
IPTS68 means the International Practical Temperature Scale which was enacted in 1968. In the meantime, IPTS68 was revised to ITS90 in 1990, thermocouples and resistance temperature detectors were reflected to New JIS1995 and to New JIS1997 respectively.
Top of this page
It is because that the room temperature is added to the setting temperature of 20 ℃ It is recommended to carry out the reference junction compensation by RJ sensor to solve this problem The use of cold junction compensator enables the reference junction to be 0℃.
In the meantime, cold junction compensation involves measuring the temperature of reference junction, calculating the temperature difference (thermoelectric power difference) from 0℃, and then carrying out compensation based on the result. When the calibrator outputs the referenced 0℃, calibrator output setting and the display of device to be calibrated do not match. This is because the terminal temperature, connecting to thermocouple is added to electromotive force from 0℃. Therefore, the calibrator is required to output the referenced terminal temperature of device under calibration. That is, measure the terminal temperature of device for calibration by means of optional RJ sensor and generate the electromotive force by reference junction compensation. Some of the calibrators can be used by selecting the built-in RJ sensor, but the accuracy may not necessarily guaranteed. For further details, please refer to instruction manual.
Top of this page
It is assumed that the DMM measuring current does not coincide with the calibrator excitation current. Please set the DMM to fixed range to meet the generated resistance value unlike the setting it to auto range. Our calibrator receives the resistance measuring current which is given by the unit to be calibrated and ouptuts(2V max) between the output terminals which is proportional to set resistance value(R) of voltage V = RXI, generating "false resistance value" of R = V/I.
Top of this page
CA11E / CA12E Handy Calibrators
The default setting of the automatic power-off feature is on. It can be turned off by removing the rear cover of the main unit and sliding dip switch no. 4 to ON.
Top of this page
The standard accessory lead cables for measurement and source are not usable to connect to a BNC cable. Instead, Yokogawa Model 369922 conversion adaptor must be used.
Top of this page
CA71 Handy Calibrator
The built-in reference junction compensation sensor is set to on by dip switch "3 INT R" to ON position. The built-in sensor will then output the reference junction compensation signal. However, for improved accuracy it is necessary to use the optional RJ sensor, B9108WA.
Top of this page
RJON display without having the external reference junction compensation sensor in use means that the internal reference junction compensation feature has been set to on by its controlling dip switch.
Top of this page
If this feature is required, the CA150 Handy Calibrator enables the changeover and setting of function/range by means of serial communication.
Top of this page
There are a number of possible causes. It may be that the equipment being calibrated uses a switching power supply. If that is so the switching noise of the equipment being calibrated is affecting the CA71 output. Grounding the equipment may lessen the noise.
Top of this page
CA150 Handy Calibrator
It is possible to set current output, voltage output and functions using the RS232C communication feature.
Top of this page
When in the MEASURE mode for pulse signal integration CPM is the count of the number of pulses per minute and CPH is the count of the number of pulses per hour.
Top of this page
Error code Err11 means that an incorrect command is being transmitted. Please note that all commands must be in capital letters.
Top of this page
Since the sweep function starts at 0V only the high limit can be set.
Top of this page
The AC adaptor works without the NiMH battery, but it is needed to charge the battery.
Top of this page
The 93026 carrying case is large enough to hold the main body plus source/measurement lead cables, terminal adapter, 6 spare batteries, fuse, AC adaptor and instruction manual. The 93027 case is open faced to hold the main unit and includes a strap and a B9108XA style case for carrying some accessories.

Attached table shows the specification comparisons among CA100 (discontinued), CA150 and CA71 for reference.