The main unit of the DC100 with optional MATH feature can perform the four arithmetic operations, integration of measured data, and computations such as detection of maximum and minimum values, in real-time. Even without the optional feature, the DC100 can compute linear scaling, difference and moving average. The results of such computations are transferred with the measured data to a PC, thus reducing system requirements on the PC and also resulting in more efficient analytical processing.
The main computing functions are as follows. (The shortest computation period differs depending on the kind of computation.)
- Standard computing functions
Linear scaling, moving average, differential calculation, pulse integration (when a pulse input module is recognized).
- Optional functions
The four arithmetic operations, logic operations, related operations, calculation of absolute and relative values, and statistical calculations (maximum, minimum, mean, and integrated values for fixed intervals)
- Moving average function
This function updates the measured value while calculating the moving average, thus effectively monitoring the trend of a varying input signal over a long period. It can also be used as a digital filter when noise components are present on the input signal. You can set the number of moving average scans by selecting a value between 2 and 64.
- Batch intergration

|