Portable Hybrid Recorder

DR130

Functions

Other standard functions

Function Description
Input Integration mode selection You can select the 50/60 Hz or 10 Hz integration mode. The 10 Hz integration mode is useful when power line noise containing both 50 Hz and 60 Hz components is superimposed on the signal. (When the 10 Hz integration mode is activated, the minimum measurement interval is 4 seconds.)
Low-pass filter You can insert a low-pass filter in the path of a signal on which noise components are superimposed. (When the hard filter is ON, the measurement interval becomes at least 3 seconds.)
Scaling The input signal is displayed and/or recorded as an industrial quantity or a physical quantity.
Burn-out When the thermocouple input goes open circuit, the indicator mores to the 100% or 0% side.
Differential computation The difference between the reference channel and measured channel is measured.
Recording Zone recording The recording area can be set freely for each channel.
Partially compressed and expanded
recording
Unimportant parts are compressed, and only necessary parts are expanded, thus enabling the
recording resolution to be increased.
Group channel trend Only channels that belong to a specified group are recorded. Switchover between groups can be done
using a remote contact.
Alarm generation channel trend Trend recording takes place only for channels that emit an alarm.
Setting Memory backup The set data is protected by a lithium battery inside the unit.
Security The unit comes with a standard password lock function, preventing mis-operation and also protecting the set data.
Alarm
(Output is optional)
Setting You can set a 4-level alarm (upper and lower limits, difference between upper and lower limits,
percentage change rising and falling limits) for each channel.
Re-breakdown re-alarm The alarm output can be refreshed when an alarm is emitted.
Hold function Once an alarm is emitted, the alarm indication and relay state are held until the operator acknowledges the alarm.
Option functions General purpose communication function (GP-IB, RS-232-C, Ethernet) and alarm output relay (10 make contacts)
DI/DO functions (recorder action control function, fail function, chart end function), etc.
Conformity to standards (based on) CSA Obtained CSA22.2 No.1010.1,
Installation category (Overvoltage category) : II, Degree of pollution : 2
UL Obtained UL3111-1 (CSA NRTL/C)
CE EMC directive EN61326
EN61000-3-2
EN61000-3-3
EN55011 Class A Group 1
Low voltage directive EN61010-1
Measurement category : II, Degree of pollution : 2
C-Tick AS/NZS 2064 Class A Group 1