Have you ever thought it might be easier to manage data files after acquisition if the data acquisition start and stop times were on the hour?
You can have data files be saved on the hour by using the MW100's file division function. This enables increased efficiency of management and application of data acquisition files.
Example settings to save data files every eight hours, starting at 00:00.
・Record setting: Direct; Condition: Full stop/Rotate; Data Length: 8 hr
・Timer Types: Absolute; Standard time: 0:00; Time interval: 8h
・EVT/ACT (Edge): Timer; ACT: Memory Save

|
|
・ You can browse measured data efficiently because data files are divided on the hour.
・ Management of data files using spreadsheet programs is made easier.
|
 |