Category Sitemap
Home / Our Businesses / Confocal Scanner Unit / Products /
CSU22
Examples of Microscope-setup of CSU22
 
Upright microscopes
micro1
 
Inverted microscopes
micro2
Note:The CSU22 can be mounted on microscopes from different manufacturers
General Specifications
 
CSU22
 
Confocal scanning method: Microlens-enhanced Nipkow disk scanning
Video synchronization: Scan-speed synchronization through NTSC or PAL composite signals
External synchronization: Scan-speed synchronization through pulse signals
Input : TTL level , 25 to 83.33 Hz
Output : TTL level , 25 to 83.33 Hz
(Corresponding to a Nipkow disk spinning speed from 1,500 to 5,000 rpm)
Excitation wavelength: Standard(CH1) : 488 nm
Optional(CH2) : 532 or 568 nm
Optional(CH3) : Multiple wavelengths (488 , 568 and 647 nm)
Contact Yokogawa for wavelength bands other than the standard wavelength
Excitation light ND filter: 0% (no light) , 10% , and 100% (through)
Laser beam input: Single-mode fiber connected with FC-connector
Fluorescence wavelengths to be observed: Standard ( CH1 ) : 520 nm or longer
Optional ( CH2 ) : 570 nm or longer, or 600 nm or longer
Optional ( CH3 ) : Multi-wavelength ( 520 to 540 nm , 590 to 620 nm , and 680 to 710 nm )
Note : For channels without filters , laser shield panel will be fixed instead of filters for laser safety.
Light path switching: Manual switching between eyepiece for direct view and C-mount camera port using the port switching knob
Operation panel: Switch for opening/closing the excitation light shutter
Switch for selecting an ND filter
Five wavelength filter combinations can be selected from three groups each consisting of three options of excitation filter(EX) , dichroic mirror(DM) , and barrier filter(BA) .
External control: RS-232C interface
Requires the use of an optional external connection interlock key and cable (specify either Windows or Macintosh).
Microscope mount: Add-on by using a direct C-mount adapter of the microscope
Operating temperature range: 15 to 30℃
Operating humidity range: 10 to 75% RH
Power consumption (main unit): 12 V DC , 1 A
Power consumption (AC adapter): Input : 100 to 240 V AC±10% , 50 or 60 Hz , 74 VA max.
Output : 12 V DC , 2.5 A max.
External dimensions (outermost): 183 (W) × 205 (L) × 245 (H) (mm)
Weight: 5 kg (main unit)
 
Note : Use of Infinity corrected microscope with high NA objective lenses (i.e.,Plan Apo) are recommended.