Category Sitemap
Home / Our Businesses / Confocal Scanner Unit / Products /
CSU-X1
General specification
  1 2 3  

basic model

Main unit *1
Main unit +
Brightfield*2
Confocal scanning method
Microlens-enhanced Nipkow disk scanning
Scanning speed

Choice:1,500~5,000rpm (Standard)
1,500~5,000rpm (High-speed)*3*4
1,500~10,000rpm (High-speed)*3*4

External synchronization
Option*4
Excitation wavelangth
405~647nm
Second port
-
Brightfield
Dichroic mirror
Option*5
Dichroic mirror switching
Manual 1CH
( Dichroic mirror block can be exchangeable)
Laser beam input
AFC-connector (Polished 8 degree)
Optical fiber
Yokogawa's standard Single-mode polarization-preserving fiber*6
Filter wheel (Emission side)
-
Emission filter
Option*5
Operation panel
Switch open / close of laser shutter
External control
-*4
Microscope mount
C -mount adapter
Operating environment
15~40℃ / 20~75% RH
Power consumption
(main unit)
24VDC 1A max.
Power consumption
(AC adapter)
Input:100~240 VAC ±10% ,50 or 60Hz 、38W max.
Output : 24VDC 1.6A max.
External dimension *7
175(W)×328.5(H)
×325.1(L) mm
258.8(W)×373(H)
×325.1(L) mm
Weight
9.4kg
11.4kg
 

*1 Supplied with a control unit ( for filter wheel ) and a filter wheel.
*2 Supplied with two control units ( one each for filter wheel and for bright field ) and a filter wheel.
*3 Option.
*4 Requires control unit for rotation speed control and external synchronization.
*5 Filters are not included ( Excitation filter , Emission filter ,Dichroic mirror ) .
Please inquire as you need.
*6 Supplied with each CSU-X1 main unit.
*7 Excluding protruding parts.

BACK NEXT
 