| Item |
Specifications |
Conditions |
| Probe length |
1.5m |
|
| Connector type |
BNC |
|
| Input resistance2 |
10 Mohm ±2% |
In conjunction with an oscilloscope with an input impedance of 1 Mohm ±1%. |
| Input capacitance2 |
Approx. 14 pF |
In conjunction with an oscilloscope with an input impedance of 1 Mohm ±1% |
| Input capacitance1 |
150 pF max. |
Probe only |
| Attenuation ratio2v |
1/10 ±2% |
In conjunction with an oscilloscope with an input impedance of 1 Mohm ±1%. |
| Bandwidth2 |
DC to 400MHz (-3 dB or less) |
Subject to change depending on type of oscilloscope used. |
| Bandwidth1 |
DC to 6MHz (-3 dB or less, typical3) |
Subject to change depending on type of oscilloscope used and measurement conditions. |
| Rise time2 |
900 ps max. |
Subject to change depending on type of oscilloscope used. |
| Rise time1 |
58 ns max. (typical3) |
Subject to change depending on type of oscilloscope used and measurement conditions. |
| Max input voltage2 |
600V(DC+AC peak) or 424 Vrms |
Frequency of the AC needs to be less than 100kHz. |
1:When a 1/1 attenuation rate is selected.
2:When a 1/10 attenuation rate is selected.
3:Typical (or average) value; not guaranteed. |