| Features (1) | Features (2) DualCapture: A Powerful Tool for Durability Test Data Analysis | Features (3) Modules | Features (4) 701280 Frequency Module (2 CH) | Features (5) 701280 Frequency Module (2 CH) | Features (5) 701275 Acceleration/Voltage Module | Features (5) Firmware | Features (6) Web Services | Features (7) DC12 V Power Supply Input Option |
|
|
All the functions and performance specs required for pulse measurement packed into a small, 2 CH module.
|
|
The module's internal high-speed CPU performs measurement, calculation, and recording in real time.
-
9 measurement modes
-
0.01 Hz-200 kHz
-
40 μs (25 kHz) update rate
-
Max 16 ch
-
Directly records and measured results
Unlike with FV converters, rescaling is not necessary with this module. Directly record measured results. Display numerical values directly using cursors and auto measurements.

1 Allowable input frequency range : 0.01 Hz 200 kHz
|
|
Real-Time Digital Filtering
|
The DL750 has 2 real-time digital filters that can be used for waveform averaging.
-
Smoothing Filter (Moving Average)
Smooth stair-step shaped waveforms.
Updating occurs every 40 μs, giving a highspeed averaging effect. Filters are set at 0.1 ms-1 sec (up to the 25000th order). Filters reduce jitter in observed waveforms, and increases resolution.
-
Pulse Average
Useful for determining the average value per rotation, or determining the number of rotations when a gear is missing teeth. Output is averaged every specified number of pulses (between 1 and 4096 pulses).
|
 |
 |
- Direct-Coupled Signal Input
- Encoder pulse input of up to ±42 V.
- Direct input to EM (electromagnetic) pickups (1:1).
- AC power up to 300 V (10:1 isolation probe 700929 required).
- Multiple Analog Inputs
Isolated input, 6 ranges, 4 input filters, internal pullup (mechanical contacts, supports open collector), AC/DC coupling, variable threshold value, 3 types of hysteresis
- Chatter Elimination (Supports Mechanical Contacts)
Time setting: 1 ms-1000 ms
- Convenient Input Presets
Presets for logic input (3 V-12 V), EM pickups, and AC power available in menus.
|
|
Supports Braking Applications
|
|
By predicting the deceleration curve and stop point, the module automatically compensates for the lack of information on encoder pulses which occur during deceleration.
-
Deceleration Prediction (1)
Automatically calculates and outputs a deceleration curve based on the interval of the last input pulse.
-
Stop Prediction (2)
If no pulses are inputted for a period of time, a stop is inferred, and output is set to 0. Up to 10 steps can be specified.
|
|
Offset Observation Function
|
An observational center (offset) is set and the area surrounding the offset can be zoomed. Useful when you wish to observe the fluctuation centering on a certain frequency.
|
Triggers can be activated on output results from all observation modes. For example, you can observe and trigger on a duty value.
|
|
|
|
|
DL Series Selection Guide Find the right one for you! |
Serial Bus Analyzer Selection Guide For I2C, CAN, and SPI analysis |
Probes for Digital Oscilloscopes Supporting a variety of measurement |