Frequently Asked Questions

DL1500 Series Portable Digital Oscilloscopes
Questions
Data Saving
Computation
Acqusition
Display
Trigger
Communication
Miscellaneous
Questions and Answers
Data Saving
Yes. When the data do not fit on one disk, a message appears prompting you to enter the next floppy disk. Simply follow the instructions to save the data to multiple floppy disks. The same procedure also applies when loading the data. These operations are also the same for the DL1520.
Top of this page
Yes, the data are retained using a lithium battery even if the power is turned OFF. This applies to the DL1520 Series and the DL1540 Series.
Top of this page
The measurement function can be used only when data are saved in the ACQ (BIN) format. In addition, when data are saved in this format, all functions including waveform zoom can be used in the same fashion as with the captured data. When P-P data are saved, the waveform can only be displayed; measurement functions and other functions cannot be used.
Top of this page
Yes, screen image data can be saved in BMP and other formats even if the built-in printer is not installed.
Top of this page
It is in the IS2 format unless the History Memory function is active, then it is in the IS1 format.
Top of this page
When using color, the file size ranges from 50 KB to 200 KB when Compression is turned ON. The size varies greatly since the compression rate changes depending on the density of the waveform. The file size is larger for waveforms that cover the entire display screen. When Compression is turned OFF, the size is approximately 300 KB.
Top of this page
This is not possible, because the internal circuit is not common between the two models. The reverse direction is not possible, either. Note that the data on the DL1540CL can be loaded by the DL1540C, but not vice versa.
Top of this page
Yes, you can select color or monochrome when saving or transferring the data.
Top of this page
There are two types of waveform data that can be loaded. Waveform data saved using P-P: Computation and X-Y display cannot be performed on P-P display waveform data that are loaded. Waveform data saved in ACQ BIN format: Computation and X-Y display are possible on this type of data.
Top of this page
Computation
It is the power spectrum defined by PS = 20 x Log(X/1Vrms)
Top of this page
If Y(V) is the converted value for X(dBV), the equation is Y = 10(X/20) (V) from X=20xlogY
Top of this page
If the FFT does not appear, check to assure that the horizontal position of the waveform is set exactly to Zero. If the horizontal position is set to a non-zero value, the FFT will not be performed.
Top of this page
The result of the computation between channels is displayed in units of divisions. By subtracting ch1 from ch2, you obtain 10 V. However, the result shows 2, a value obtained after dividing 10 V by the V/div value (10/5).
Top of this page
Acquisition
The averaging function supported by the DL1500 Series is exponential averaging. The models that support linear averaging (simple averaging) are the DL4000 Series, DL2700, and DL708.
Top of this page
The minimum dead time on the DL1540 and DL1540L is approximately 200 µs. However, this value varies depending on the T/Div setting, the number of channels, and the data length. Consider the minimum value to be the value obtained when the memory length is set to 1 KW and the T/Div is set to 1 µs/Div. For other settings, the dead time may even be a few milliseconds. Therefore, keep in mind that the value indicated above is a minimum value.
Top of this page
Display
The FFT is computed based on the ACQ data. In order to perform the FFT over 1 k points, the ACQ data are decimated (reduced) from 10 kW to 1 kW. At this point the sampling rate is equivalent to 100 ks/s (1/10 of 1 MS/s). Therefore, the sampling rate display is turned OFF. For example, to perform an FFT over the data sampled at 1 MS/s when the memory length is 10 kW, the data are decimated from 10 kW down to 1 kW. This means that the sampling rate is equivalent to 100 kS/s at this point. Therefore, to avoid misinterpretation, the sampling rate display is turned OFF. In this case, the actual sampling rate is 1 MS/s, which means that the frequency bandwidth of the time axis waveform is 500 kHz. However, because the sampling rate is reduced to 1/10 as explained above, the frequency bandwidth of the FFT is 50 kHz.
Top of this page
It is fixed. 1ch: yellow, 2ch: green, 3ch: pink, 4ch: blue.
Top of this page
Yes, the color of the snapshot waveform is white.
Top of this page
The manufacturer of our LCDs, has given us data indicating that the luminance drops to half after 25000 hours.
Top of this page
Trigger
The width level of the WINDOW trigger cannot be set to a value less than the voltage corresponding to 1 division at the specified voltage sensitivity. For example, if the voltage sensitivity is set to 500 V/div, then the width level cannot be set below 500 V. This limitation is due to the fact that the trigger sensitivity of the DL1500 Series is 1 divp-p.
Top of this page
Yes, there is no restriction on specifying or using the two functions simultaneously.
Top of this page
Communication
This function is available on firmware version 1.10 and higher. To check the version, press the MISC key on the front panel. Then select "Information" using the soft keys. This function is available on all versions of the DL1540L.
Top of this page
The menu screen and messages are cleared when commands such as "start" and "cursor on" are issued.
Top of this page
Only P-P data can be transferred in the started condition. The instrument must be stopped to transfer ACQ data.
Top of this page
Up to 10 seconds. Normally, it is six or seven seconds, however this will vary depending on the performance of the PC and the GP-IB board. On the DL1540C, this may take longer than 10 s, because the screen image data contain color information. Assure that the time-out setting for the PC being used is set long enough to accomodate the required transfer time..
Top of this page
Miscellaneous
Error code 915 is the error message "Can't detect listener." This is described in the Communication Interface User's Manual. This error occurs when the hard copy destination is not set to the "printer", but to "plotter" without connecting external devices.
Top of this page
It is 1 to 2 Hz when all measurement items are specified on Trace1 through Trace4. It is 5 to 6 Hz when only a few measurement items are specified on Trace1.
Top of this page
Approximately 500 ms when performing zone determination and Approximately 700 ms when performing parameter determination.
Top of this page
No. The power supply section of the DL1540, DL1540L, DL1540C, and DL1540CL automatically switches between 100-120 VAC and 220-240 VAC. The power supply section of the DL1520 and DL1520L is specified when you order the product. It does not have a power selector switch, and cannot be modified after the product is delivered.
Top of this page
Yes. The part number is A9080ZG.
Top of this page
The NO-GO output is controlled by software. Thus, there may be fluctuations in the processing time depending on the sensing conditions. Since the signal remains low until the next determination can be accepted, this means that the processing time may fluctuate at the software level depending on the conditions. In any case, the minimum pulse width of 10 ms is guaranteed.
Top of this page
Half-pitch 50-pin (pin type).
Top of this page
No, it can't. As the DL1540C, DL1540CL use the color TFT LCD screen, output to the monitor is supported by the VGA output function (optional).
Top of this page
To use the isolated input unit, the firmware version of the DL1540 must be 1.10 or higher. Please check the version number. If the version number is lower than 1.10, the firmware must be upgraded.
Top of this page
You can display a comment at the top section of the waveform display frame. In this case, up to 25 characters can be entered. Up to 25 characters are also displayed. However, note that if you attempt to edit a preexisting comment, only the first nine characters are displayed. In this case, you must reenter characters after the ninth character.