Category Sitemap
Home / Our Businesses / Test & Measurement / Support / FAQs /
Frequently Asked Questions

WT2000 Digital Power Meters
Questions
Questions and Answers
12 bits.
Top of this page
A BNC connector. The output terminal on the Clamp Probe (Model 751550) is a banana plug. Use the Banana-to-BNC Conversion Adapter (Model 366921, sold separately) to connect to the external shunt input terminal.
Top of this page
Output function: HAr
When the default output item is set to 1, the printout time for up to the 50th order is about 116 seconds. The printout time for up to the 25th order is about 75 seconds.

Output function: HAr
When the default output item is set to 2, the printout time for up to the 50th order is about 41 seconds. The printout time for up to the 25th order is about 27 seconds.
Top of this page
In order to analyze the harmonics by the PLL synchronization, the maximum possible window width is automatically determined depending on the frequency of the fundamental wave of the input signal (fundamental frequency f).
  • When the fundamental frequency f of the input signal is 40 =/< f < 70 Hz, the width cannot be set higher than 16.
  • The following are the settings possible given different fundamental frequencies f:
    10 =/< f < 20 Hz: 1 to 4
    20 =/< f < 40 Hz: 1 to 8
    40 =/< f < 70 Hz: 1 to 16
    70 =/< f < 130 Hz: 1 to 32
    130 =/< f < 250 Hz: 1 to 64
    250 =/< f =/< 440 Hz: 1 to 128
Top of this page
Yes, this is possible. The following are the range standard special order specifications and corresponding models.
  • Current range 1/10 model: WT1010/WT1030/WT1030M/WT2010/WT2030
  • Current range 100 A model: WT1010/WT1030/WT1030M