The GS820 allows the generation trigger source and measurement trigger source to be set separately.
There are two types of constant period timers and an external signal input that can be used for the generation trigger source.
In addition to these sources, source change point and sweep end point can be used for the measurement trigger source.
Because the source trigger and measurement trigger can be set separately and also separately for each channel,
source and measurement under various connection conditions and timing combinations can be accommodated.
There is also an auxiliary trigger that can be activated using an external signal or a program event.
The source delay, measure delay, and integration time can be set separately for each channel.