In addition to the preset sweep functions described above,
the GS820 is equipped with a programmable sweep function
that allows the user to defi ne the sweep pattern.
A user can create or edit arbitrary waveform data (CSV format) of up to 100,000 points
using a spreadsheet or text editor.
The GS820 is also of capable of sweeping the timing and control parameters
in addition to the source level.
This allows a control sequence that is synchronized to the waveform generation timing.
The sweep program can be changed
at a minimum of 100-µs intervals*¹ in programmable sweep mode.
*1: Minimum program cycle
100 µs: If the measurement function is OFF
500 µs: If the measurement function is ON, two channels are used, and auto zero is OFF
The values do not include the hardware operation time corresponding to the control parameters.