Tunable laser Source

AQ2200-136
Features
  1 2  
Example Measurement Application
1. Swept measurement synchronized with optical spectrum analyzer
Example Measurement Application
2. Swept measurement system
Example Measurement Application
3. Wavelength meter monitor
Specifications
(incl. Laser safety information)
3. Wavelength meter monitor (standard function)

  • When combined with the AQ6141 wavelength meter, the AQ2200-136 can be used as ahigh-accuracy wideband variable wavelength light source, with wavelength accuracy of ±3 pm*. (Wavelengths are automatically corrected without an external PC).
    * This accuracy value does not include the wavelength accuracy of the wavelength meter.

  • The internal wavelength meter monitor port can be used to perform wavelength correc-tions without switching measurement ports.
  • Wavelength meter monitor
    Parameter Product specificationRemarks
    Wavelength varying range 1440 to 1640 nm
    Wavelength resolution 1 pm
    Absolute wavelength accuracy ≤±100 pm1
    Relative wavelength accuracy ≤±50 pm1
    Wavelength stability ±5 pm/h2, 4, 5
    Spectral line width NARROW ≤1 MHz2, 4, 8
    WIDE ≥50 MHz1, 2, 4, 8
    Optical output Maximum output wavelength ≥±7 dBm1, 10
    1520 to 1610 nm ≥±5 dBm
    Total wavelength range ≥-8 dBm
    Optical output stability 5 minutes ≤±0.01 dB1, 3, 5, 6, 10
    1 hour ≤±0.05 dB
    24 hours ≤±0.1 dB
    Optical output repeatability ≤±0.04 dB1, 3, 5, 6, 10
    Monitor output Maximum optical output ≤±5 dBm9
    Minimum optical output ≥-25 dBm
    Accept optical fiber SM (ITU-T G.652)
    Accepted optical connector FC/Angled PC7

    Laser safety information
    INVISIBLE LASER RADIATION

    • This laser light source is catego-rized as an IEC60825-1:2001 Class 1 M product.
    • This product complies with 21CFR1040.10, except forexemptions resulting from itscompliance with Laser Notice No.50, dated May 27, 2001.
    • Viewing laser output at a distance of 100 mm or less using someoptical measuring devices (loupes, magnifying glasses, tele-scopes, etc.) may result in eye injury.

    Unless otherwise noted, the specifications represent values obtained after calibration performed at least one hour OPTON, under reference operatingconditions * with 2 m SMF output point.
    * Reference operating conditions: Ambient temperature of 23±2ºC, temperature change rates of 1ºC per 10 minutes and 3ºC per hour, and humidity of 50±5%

    1 At maximum optical output
    2 Wavelength: 1590 nm
    3 Wide line width
    4 Narrow line width
    5 Constant temperature
    6 Wavelength: 1460 to 1620 nm
    7 SEIKOH GIKEN Angled/PC or conforming product (step type)
    8 At minimum optical output
    9 Using MONITOR connector
    10 Using OUTPUT connector