Category Sitemap
Home / Our Businesses / Test & Measurement / Products / Optical Fiber Test Equipment /
Bit Error Rate Tester

AQ2200-601
Features
  1 2  
FeaturesMainframe ScreenshotExample ApplicationsSpecifications

Optical device characteristics evaluation test

Optical device characteristics evaluation test

The AQ2200-601 can be used to test and evaluate the characteristics of optical devices such as optical modulators and transmitter optical subassemblies (TOSA). Waveform characteristics can be observed using instruments such as an oscilloscope, optical spectrum analyzer, and optical power meter while varying the amplitude, offset, and crosspoint of the data input signal being input to the modulator driver.
  • Rise and fall times: Tr/Tf, Crosspoint position
  • Overshoot percentage
  • Jitter measurements: Peak to peak and RMS
  • Extinction ratio
  • Pulse mask test
  • Output power (peak, average), light wavelength

XFP module characteristics evaluation test

XFP module characteristics evaluation test

The XFP module is an MSA which is an optical transceiver containing an optical transmitter and an optical receiver. XFP technology is very flexible, with support for STM-64/OC-192, 10 G Fibre Channel, and 10 G Ethernet. The XFP module requires the following tests:
  • Jitter test
  • Optical reception sensitivity test
  • Optical output eye mask test
With AQ2200 Series modules, it is possible to install the BERT, optical modulator, optical receiver, optical ATTN, and other equipment in a 19-inch rack to form an extremely compact test system. The AQ2200 series can also handle optical receiver sensitivity tests.
PPG and ED
Item Product specifications
Internal clock Internal frequency variation Variable frequency range: 9.95 GHz to 11.32 GHz
Minimum setting increment: 1 kHz
External reference Frequency 1/64 or 1/16 bit rate
Input level 0.4 to 1.0 Vp-p: 50 Ω AC
External clock input Frequency 1/1 bit rate
Input level 0.4 to 1.0 Vp-p: 50 Ω AC
Trigger output Clock trigger 1/64 or 1/16 clock output frequency
Pattern triger PRBS, PROGRAM
Output level 0.6 Vp-p ±0.3 V

PPG only
Item Product specifications
Data output Bit rate 9.95 Gbit/s to 11.32 Gbit/s
Output level 0.50 to 2.00 Vp-p (in steps of 10 mV)
Offset voltage -2 V to +3 V (in steps of 10 mV)
Crosspoint variation 30% to 70% (in steps of 1%)
Connector 3.5 mm female
50 Ω AC termination or DC termination
Clock output output level 0.6 Vp-p (AC coupled)
Offset -2 V to +3 V (in steps of 10 mV)
Connector SMA female
50 Ω AC termination or DC termination
Transmit patterns PRBS 7, 9, 10, 11, 15, 23 or 31
PROGRAM (1) 16 to 256 bits (in steps of 1 bit)
(2) Option for 256 - 67, 108, 864 bits (in steps of 128 bits)

ED only
Item Product specifications
DATA IN (CDR) Bit rate 9.95 Gbit/s to 11.32 Gbit/s
Input level range 0.1 to 0.7 Vp-p
Input threshold value variation ±0.35 V (in steps of 1 mV)
Connector 3.5 mm female: 50 Ω AC coupling
DATA IN Bit rate 9.95 Gbit/s to 10.71 Gbit/s
Input level range 0.1 to 0.6 Vp-p
Input value variation ±0.3 V (in steps of 1mV)
Connector 3.5 mm female: 50 Ω AC coupling
Clock input Frequency Frequency with same bit rate synchronized to data input
Input level 0.2 to 0.6 Vp-p
Connector SMA female: 50 Ω AC coupling
Receiving patterns Same as PRBS and PROGRAM sending patterns
Viewing PDF files requires Adobe Reader.
GLOBAL  
YOKOGAWA

Powered by Ultraseek
·  AQ2200-601
  1 2