- Ultra high Wavelength Accuracy
±10 pm (1520 to 1580 nm)
±20 pm (1450 to 1520 nm)
±20 pm (1580 to 1620 nm)
±50 nm ( 600 to 1700 nm)
The above specs are after calibration with a built-in source
- High Wavelength Linearity
±10 pm (1520 to 1580 nm)
±20 pm (1450 to 1520 nm)
±20 pm (1580 to 1620 nm)
The above specs are after calibration with a built-in source
- High Wavelength Resolution
Provides maximum wavelength resolution at 10 pm
- High Wavelength Resolution Accuracy
Provides excellent wavelength resolution accuracy within ±2%
- Best Optical Performance
Wide close-in dynamic range:
70 dB @ ±0.2 nm
Fast sweep and quick response
- Versatile Analysis Functions
Comes with zooming and overview windows
- Low Polarization Dependency
Measurements such as gain of the optical amplifier can proceed accurately because polarization dependency is suppressed as low as ±0.05 dB.
- High Power Measurement
Max input power: +23 dBm per channel; Max safe input power: +27 dBm
- Multiple Communication Interface
GP-IB, LAN, printer, and oters.
- Easy Operation with Mouse and Keyboard
- Large Data Storage and Fast Data Transfer (FTP)
|