The AQ6375 comes with the built-in calibrator for wavelength calibration and
optical alignment adjustment of the monochromator in order to maintain accurate
measurements. The calibration and adjustment are automatically performed and
completed within a couple of minutes.
Wavelength Calibration
Calibrates with the designate absorption line of Acetylene gas.
Optical Alignment Adjustment
Corrects mechanical movements inside the monochromator caused by shock and
vibration.
Note. the wavelength can also be calibrated with an external light source.