The integration method used in the WT2000 series enables the instrument to continuously acquire data by sampling the input signal at a high rate of about 110 kHz. As a result, the instrument can measure abrupt changes in input value, which was difficult to realize with conventional measuring instruments, thus enabling integrated values to be measured with very little loss of data. (There is virtually no delay or pause period due to the response of the filter.) With YOKOGAWA's conventional models, the resolution of the displayed value resulting from an integration operation is fixed by the maximum integration time, hence for short period integration, the bottom digit was not displayed and the resolution was insufficient. The WT 2000 series uses an "auto resolution method" in which the display resolution automatically changes along with the lapse of the integration time, ensuring the maximum resolution even for short period measurement.
|