Are there any special requirements for the replaceable electrode structure model of AXF?

​Per the general specifications (GS), units must be 1-12", PFA or Polyurethane Rubber liner, and the electrode material must be 316L stainless steel only. Also, replaceable electrodes must be ordered with the initial flowmeter because electrodes are manufactured and fit to a particular flowmeter. In addition, a special Yokogawa tool is required to extract and replace electrodes. Please contact a local sales representative or factory prior to ordering.

Related Products & Solutions

  • ADMAG AXF

    The AXF magnetic flow meter is a sophisticated product with outstanding reliability and ease of operation, developed on the basis of decades of field experience. Based on FOUNDATION™ fieldbus specifications, AXF Fieldbus Magnetic Flow meter models offer more flexible instrumentation.

    See More