Below are 38 results from your search.

How do I change Temperature units in my GX ?

  See attached file for explanation. ​  ...

Can I add additional I/O to my GX in the field ?

  Yes, The GX series recorders are modular so additional I/O can be added or removed in the field.     See file attached for explanation. ​...

Can I eliminate process error at my GX ?

Yes, The GX and GP series recorders have a built in calibration correction function. The calibration correction feature is used to correct the input value using segments, and the result is used as a measured value. This ...

How do I view data files from my GX20?

Data files must be viewed using the universal viewer software. The software is available for download from the smartdac plus website @          www.smartdacplus.com​...

What registers are used for modbus communications in GX/GP series ?

The registers have been changed in the GX/GP series and different registers are used than in the DX Advanced and DX100 series. For example, holding registers in the DX Advanced and the DX100 series were 40000 registers s...

I keep getting an error message on my GX that says: Scale image not loaded. How do I get rid of this error ?

  The scale image setting in the GX/GP data aquisition products is on by default when the units are shipped from the factory. This error message will keep appearing until this setting has been turned off.   To turn off...

What is the correct time zone setting for my GX recorder?

  The time zone setting is the difference between your time zone and Greenwich mean time.The default time zone setting for the GX/GP series units is set to 9:00 hrs when shipped from the factory. The correct time zone s...

How many server devices does GX support when using as modbus client?

  There are 2 different memory options for the GX20 series.   Standard memory is 500 MB which will support up to 16 modbus server devices and 100 client command lines.   The GX20 is also available with the large memor...

Can I start and stop recording remotely on my GX20 recorder?

  Yes, recording be started and stopped remotely. Using the web browser function you can start/stop recording, start/stop math computation, and acknowledge alarms. ​  ...

What type of data files does the GX/GP units create ?

  See list below for all GX/GP file types and extensions:   Display Data files: .GDS - Binary (undisclosed) .GTD - Text format (TSV)   Evenat data files: .GEV - Binary (undisclosed) .GTE - Text format (TSV)   Report ...

How many Admin and User logins are available on the GX/GP units ?

  There are 50 logins available on the Smartdac plus series. There is no specific number of Admin logins and user logins that can be set.   The only limitation is that at least 1 of the 50 available logins must be an A...

How many groups are available and how many channels per group can be set on my GX/GP unit?

  There are 30 groups available on the GX10/GP10 units and 10 channels can be assigned to each group.   There are 50 groups available on the GX20/GP20 units and 20 channels can be assigned to each group. ​  ...

What is the part number and price for the GX IQ/OQ protocol?

​ The part number is 773230, and for price please contact your local sales rep for a quote.  ...

1. What Option Do I need on a GX10/GX20 to meet the specifications in a cGMP (21 CFR PART11) environment?

​The option on the GX10/GX20 that is needed in the cGMP environment is the ADVANCED SECURITY FUNCTION (/AS):   Security and electronic record/signature functionshave been added that are compliant with the USA’sFDA t...

2. What Option Do I need on a GX10/GX20 to be able to build custom screens to be displayed on the GX?

​The option on a GX10/GX20 that is needed to program custom screens on the GX10/GX20 is the Custom Display Option (/CG) and the DXA170 DAQStudio software:   Using DXA170 DAQStudio, screen creationsoftware, a custom sc...

3. What are the physical communication connections on the GX10/GX20 Recorder?

​There are three communications ports on the GX10/GX20 recorder: EIA RS-232 (/C2), EIA RS-422/485 (/C3), and a single RJ45 Ethernet connection. ( Note: the RJ45 Ethernet option is standard on the GX10/GX20 recorders, b...

4. What are the available communication protocols to the GX10/GX20 products?

​ There are several communication protocols in the GX product family. Most are listed below, but for more details, please down load the General Specifications sheet ( GS 04L51B01-01EN ). A link is provided below.   ...

What size display screens are on the GX10/GX20 Products?

GX10: 5.7-inch TFT color LCD (640 × 480 dots) GX20: 12.1-inch TFT color LCD (800 × 600 dots)  ...

How many characters will the GX/GP units Batch number field hold

Batch number (up to 32 characters) + lot number (up to 8 digits) ​...

Can the GX/GP units be set up to email report data files

  Yes,   Email can be sent by events as below:   - Alarm occurring/alarm canceling- Recover from power failure- Report data generating- Storage medium error, FTP client function error- Specified time period   ​  ...

What is the power supplu rating for the GX/GP units

Rated supply voltage: 100 to 240 VAC Allowable power supply voltage range: 90 to 132,180 to 264 VAC Rated power supply frequency: 50/60 Hz ​...

What is the rated power consumption for the GX/GP units

Supply Voltage   100 VAC            GX10: 16 VA with backlight off                          GX20: 28 VA with backlight off                            GX10: 2...

What is the GX/GP series recorders 24 V DC/AC Power supply rating

24 V DC/AC POWER SUPPLY (/P1)• Rated power supply: 24 V DC/24 V AC (50/60Hz)• Allowable power supply voltage range: 21.6 to26.4 V DC/AC• Insulation resistance: 20 MΩ or greater at 500 VDC between power terminal an...

What is the contact rating for the GX90XD Digital input module

• Contact rating: 12 V DC, 20 mA or more• Input resistance: Approx. 1 kΩ• Allowable input voltage: 10 V     Voltage-free contact input: Contact resistance in ON state: 200 Ω or lessContact resistance in OFF sta...

What is the rated load voltage and current rating for the GX90YD digital output module

• Rated load voltage: 24 V DC or 250 V AC or less • Rated load current: 3 A (DC)/3 A (AC), resistance load,  each channel • Min. load voltage/current: 5 V DC   • Withstand voltageBetween the output terminals an...

What is the required PC Operating environment to support Smartdac + Standard Universal viewer and Web app / Hardware configurator software

Operating environmentWindows XP Home Edition SP3 (Except for Professionalx64 Edition) Professional SP3 (Except for Professionalx64 Edition) Windows Vista Home Premium SP2 (Except for 64-bitsEditions) Business SP2 (Except...

What are the external and panel cutout Dimensions of the GX10 product?

Please download the service drawing listed below.  ...

What are the external and panel cutout Dimensions of the GX20 products?

​Please download the service drawing listed below.  ...

What size SD card does the GM10 use

The GM10 will work with up to a 32 GB SD card. The brands we recommend are sandisk and transcend. ​...

What communication protocols are available for GM10

The GM10 is available with several different communication protocols.   Available protocols are listed below:   Ethernet IP Modbus TCP Modbus RTU (RS232C, RS422/485) Bluetooth USB ​...

How much internal memory and how many I/O Channels are available with GM10

Internal memory and max. I/O channels GM10-1 500 MB internal memory   Single-unit configuration 0 to 100 -  I/O ChannelsMulti-unit configuration 0 to 100 -  I/O Channels   GM10-2 1.2 GB internal memory   Single-un...

What is the measurement accuracy for the GX90XA analog input module

The measuring accuracies noted in the general specifications have a margin of error that takes into account the product's components and the equipment used for adjustment and testing. However, the actual values calculate...

How many channels can be specified on 1 GM System

Support measurement of up to 420 ch (actual input) by expanding channels across multiple units (multi-unit configuration) Expand up to 420 ch by using the GX90EX expansion module.   (GM10-2) On the GM10-2 large capacity...

Can I make setting changes and monitor the GM with a tablet?

Yes,   Because GM10 supports Bluetooth (optional code /C8)You can enter settings or monitor from a tablet without ever bringing a PC to the site. ​...

What Security functions are available on GX/GP models

Security functions:    Key lock function, login function Clock functions: With calendar function, accuracy:±5ppm(0 to 50°C) LCD saver function ​...

How do I upgrade the firmware on my GM10 ?

See document attached. ​...

Can a value be sent using modbus communications to start/stop recording ?

Yes,   A value of 1 would be written to holding register 409001 to start recording and a value of 2 would be written to register 409001 to stop recording.   For more examples refer to GX/GP user's manual. ​...