Cross Reference - Recorder Model # vs Name

​ Many of the current and obsolete recorder models have a Model Number and a Model Name. These are sometimes confusing and difficult to find. Attached is a cross reference.  ...

What is the shelf life for Yokogawa Pen and ribbon consumables ?

  These items are manufactured by a sister Yokogawa affiliate company, Kokusai Chart Corporation, located in Japan.All pen and ribbon products are packaged in sealed metalized envelopes specifically designed to prevent...

Can I do a totalizer function with a UR10000 series chart recorder?

  Yes, A totalizer function can be programmed in the UR series chart recorders. See attached document for setting up totalizer. ​  ...

How do I change Temp units from Deg C to Deg F on my UR20000 series recorder ?

  To change temp units from Deg C to Deg F, see instructions attached. ​  ...

I am using a UR10000 chart recorder and want a periodic printout every hr. Is this possible and if so, how do I set this up ?

  Yes, It is possible to get a periodic printout every hr as long as the chart speed is fast enough. See procedure attached for set up instructions. ​  ...

SR1000 Accuracy specification when using scaling input

See attached ​...

Does the uR10000 and or uR20000 have an option for storing electronic data?

​Yes, the uR series recorders now have the EM1 option. /EM1: SD Memory CardSD memory cards can be used as external storagemedia.Measured data can be saved and replayed (only fordot model), and setting parameters can b...

Can I do a offset and or a PV input bias On a uR10000 or a uR20000 recorder?

​Yes, you can do this with a option called CC1.    / CC1: Calibration correctionCorrects the mesurement value of each channelusing segment linearizer approximation.Number of segment points: 2 to 16Setting method: Bi...

I need a list of available options for the uR10000?

​Please see page 11 in the GS sheet listed in the download below. ...

I need a list of available options for the uR20000?

​Please see page 11 in the GS sheet listed in the download listed below. ...

What is the SR10000 recorder?

​The SR10000 is a compact recorder with a recording width of 100mm. The model family consists of 1, 2, 3, 4 pen and a 6 dot models. Pen model realizes continuous recording for each channel, whereas the 6-dot model real...

What types of inputs are supported by the SR10000 recorder?

​Inputs: Inputs: DCV: Direct Current Voltage input20mV to 50V, 1-5V range.TC: Thermo couple.RTD: Resistance Temperature Detector.DI: Digital Input (contact or DC Voltage, TTLlevel).DCA: Direct Current Input (using exte...

What is the measurement interval of the SR10000 recorder?

Measurement Interval:Pen model: 125ms / channelDot printing model:1s / 6 dot (AD integration time is 20ms or 16.7ms)2.5s / 6 dot (AD integration time is 100ms)   ...

What are the safety and EMC standards for the SR10000 recorder?

​Safety and EMC standards:CSACSA22.2 No.61010-1 (NRTL/C*) installation category II,measurement category II pollution degree 2* For marking that includes NRTL, a mark with “US”(USA) printed on the right side of the ...

What are the normal operating conditions of the SR10000 recorder?

​Normal Operating Conditions:Power voltage: 90 to 132, 180 to 264VACPower frequency: 50Hz ± 2%, 60Hz ± 2%Ambient temperature: 0 to 50°CAmbient humidity: 20 to 80% RH (at 5 to 40°C)Vibration: 10 to 60Hz, 0.2m/s2 or ...

What software is used by the SR10000 recorder?

​ With Ethernet (/C7), RS-422A/485 (/C3), or Interface unit,SR10000 setting can be configured with the RXA10 Configuration Software  ...

What is the standard paper, the ink ribbon, and the ink pens on a SR10000 recorder?

The standard paper part number is B9565AW. The ribbon part number for the dot model is B9901AX. The Red pen for channel one on a pen model is B9902AM. The Green pen for channel two on a pen model is B9902AN. The Blue pen...

Can I get external alarm relays on the SR10000 recorder?

Yes. You can get it with the A1 option that has two output relays or A2 that has 4 output relays, or the A3 option that has 6 output relays.  ...

What type of display is on the SR10000 recorder?

The SR10000 has a 7 segment, 2 + 5 digits LED display.  ...

If you need more details on the SR10000 paper recorder, Please download the GS sheet.

​Download the GS sheet listed below. ...

What is the relation between chart speed and printing intervals of periodic printouts?

Relation between chart speed and printing intervals ofperiodic printouts (For AUTO interval setting): (Pen-model)                 Printing Interval of Periodic Printout Chart Speed 5 to 9mm/h      ...

What are the functions available when using the remote option on the UR series recorders

5 are selectable from the below mentioned remotecontrols. Number of settings Recording start / stop Chart speed change Message printout start Manual printout start Alarm ACK Time adjustment (Adjusting the time to a pres...

What application software is available for the UR10000/UR20000 series chart recorders?

APPLICATION SOFTWAREWith Ethernet (/C7), RS-422A/485 (/C3), or Interfaceunit, μR10000 setting can be configured. • RXA10 Configuration SoftwareSystem requirements: OS: Windows 2000/XPProcessor: Pentium III/600 MHz or ...

What types of measurements are recognized by the UR10000/UR20000 series chart recorders?

Model1, 2, 3, and 4 pen, 6 dot-model. InputInputs: DCV: Direct Current Voltage input20mV to 50V, 1-5V range. TC: Thermo couple. RTD: Resistance Temperature Detector. DI: Digital Input (contact or DC Voltage,TTL level). D...

Whatis the measurement interval and Integration time for the UR10000 series chart recorders?

Measurement Interval:Pen model: 125ms / channelDot printing model: 1s / 6 dot, 2.5s / 12-24 dot2.5s / 6 dot, 5s / 12 dot, 10s / 18-24 dot (AD integrationtime is 100ms)A / D Integration Time:AUTO / FIX selectableAUTO: 20m...

What is the recording period for the UR10000 series chart recorders?

Recording Period: Pen model:Continuous for each channel.   Dot printing model*:Max. 6 channel / 10sec(the shortest recordingperiod) 7 to 12 channels / 15sec(the shortestrecording period) 13 to 18 channels / 20sec(the sh...