Question
CW120/CW121 Clamp-on Power Meters
1.  There are three available cases for the CW120 series, how do they differ?

2.  What are the recommended CF data saving cards for the CW120 Series clamp-on meters?

3.  Can the CW120 series be operated by batteries?

4.  What does error message "Err.300" mean?

5.  What does error message "Err.105" mean?

6.  Integral power display unit is read MWh, but it seems too big.
What should we do ?

7.  Icon of MEM is shown in the display. What does it mean ?

8.  In an integral power measurement, continuous integration leads to overflow.
What is the integral power display limit ?

9.  The measured current values are radically different than expected values.

10.  Data is shown in Mocrosoft Excel for only up to the middle of the first day for a data file containing a volume for one week of 1-second interval readings. Why?

11.  The setting is transferred to the main unit by means of parameter setting tool, Toolbox, but the display shows "failed".

12.  Is it possible to monitor multiple CW120/CW121s using a Pc? If it is possible how would it be configured?

13.  Is there any simple setting way of loading the same configuration to a multiple number of CW120/CW121 units?

14.  What is the preferred communication cable for RS-485 ?

15.  I don't have a RS232 port on my compurter. Can I use a RS232 to USB converter?

16.  Please advise how to use the command to download the data by means of PC link communication.
Please advise how to handle 2 registers being returned.

17.  Are Models 96034/96035 Clamp-on probes for 3000A usable for CW120/121 Clamp-on Power Meters ?
CW240 Clamp-on Power Meters
Answer
CW120/CW121 Clamp-on Power Meters
The following describes the differences between the three CW120 series cases.
a1 )The 93023 case holds only the main unit of the CW120 or CW121.
      It does not accommodate any accessories, but includes a magnet and stand.
a2 )The 93022 carrying case holds the CW120 series main unit along with
      accessories including current clamps and voltage probes.
a3 )The 93024 case is meant to be a portable carrying case for a CW120 series
      clamp-on meter and its probes and clamps.
Top of this page
Recommended CF cards for the CW120 series are Yokogawa number 97034, 256 MB and number 97035, 512 MB.
Top of this page
No, the CW120 series clamp-ons are not battery operated. If there is no nearby AC 100 V ~ 240 V +/- 10 %, 50/60 Hz outlet available consider using the model 98030 power cable which derives power for the measurement circuit from the system being measured.
Top of this page
Error 300 is displayed when attempting to change either the measuring range or phase wiring while the integrated power consumption or elasped time have not been reset to zero. These changes cannot be done to avoid discontinuity of data. To reset the integral values to zero depress the ESC key for more than 3 seconds before trying to make these changes.
Top of this page
Error 105 means the PC card is not properly formatted. Confirm that you are using a PC card confirmed for use with the CW120 series by YMI and that the card has been properly formatted.
Top of this page
It seems that the measuring integral power consumption and display unit do not match each other. Display unit and decimal point position can be changed as required.Default is set that the decimal point position and display unit are automatically selected by the measuring conditions of voltage range, current range, wiring setting, CT ratio and/or VT ratio.In order to obtain the most appropriate display unit, select the menu and change the display unit, kWh for instance, and decimal point position to a proper display.For further details, please refer to instruction manual, Clause 7.2.17 Setting Decimal Point Position and Unit of Electric Energy (DIGIT).
Top of this page
It means that the measuring data to be saved to PC card is saved in the back-up memory of CW main unit for some reason. (When no data has been saved in the back-up memory, "MEM" is not displayed).If such back-up data is to be saved continuously, reslot the PC card and move to MEM menu and select copy and clear in back-up memory,
Top of this page
Integrating display limit is 99,999 GWh.
Top of this page
Please verify that the clamp model number in clamp setting matches the model number of clamp in use.Set the current range on main unit to match the maximum current range of the clamp probe used.
*Model 96033: 50A
*Model 96030: 200A
*Model 96031: 500A
*Model 96032: 1000A
Top of this page
It is because that the data exceeds the Excel worksheet line limit of 65, 536 lines. Please split data files into less than 65,536 lines by editor or the like.
(*) Registered trademark of Microsoft Corporation, USA
Top of this page
It is assumed that both integral power consumption and integral elapsed time are not zero. Reset the integrated value in the setting display or depress the ESC key for more than 3 seconds to make the integrated value to zero
Top of this page
RS-485 allows to connect up to 31 units of CW120/CW121 with one communication line for remote monitoring.A sample configuration would consist of a PC with RS-232, RS-232/485 converter and PC software.A RS-232C/RS-485 converter is required to connect to CW120/CW121 (RS-485 communication spec) to the RS-232 port on PC.
Top of this page
The CF card saves Configuration data as well as measuring data. Therefore, save the Configuration file CW120/CW121 to the CF card. Then insert CF card into each unit and upload the Configuration file to each unit.
Top of this page
AWG24 shielded twisted pair cable is preferred.
Top of this page
It is possible to connect by combining commercially available RS-232/USB converter and CW120/121cable (model: 91011).
Top of this page
Two (2) pieces of registers save the voltage and current data as explained in register map exclusively used for CW120/CW121. Unlike reading out them individually and add, continuous reading out is available by specifying the first register and key word to be read out as explained in WRD command. For further details, please refer to user's manual communication functions Clause 5.4.7 WRD command on P. 5-12.
Top of this page
    Yes. Even though there are no descriptions in the instruction manual
it is available in the following manner as shown in the table below.

Clamp
Setting Procedures for CW120/121
Accuracy
96034
* Clamp setting: set to 200A
* CT setting:
In case of 1000A clamp range, set CT ratio to 5.

In case of 2000A clamp range, set CT ratio to 10.

In case of 3000A clamp range, set CT ratio to 15.

* Measurement accuracy
20 - 200A
(x CT ratio in every range), at power factor 1(clamp included)
Current/active power:
±(1.2% rdg + 0.8% rng)


* Power factor effects
20 - 200A
(x CT ratio in every range),
at 45Hz≦f≦65Hz
Active power:
±2.0% rng cosФ=±0.5
(against power factor 1)
Reactive power:
±2.0% rng sinФ=±0.5
(against reactive factor 1)

96035
* Clamp setting: set to 200A.
* CT setting:
In case of 300A clamp range, set CT ratio to 1.5.

In case of 3000A clamp range, set CT ratio to 15.

※ rdg:reading, rng:range

Please also be noted that the Mdoel 96036 Clamp-on probe for 2A is not used for CW120/121.
Top of this page
CW240 Clamp-on Power Meters
Recommended CF cards for CW120/240 are Yokogawa number 97034, 256 MB and number 97035, 512MB. Operation has been successfully checked with the following commercially available cards with more memory, but operation

SanDisk SDCFX3-002G-J21A (Extreme III CompactFlash 2GB)
SanDisk SDCFB-1024 (1GB)
SanDisk SDCFB-2048 (2GB)

* No other files are presented in PC card.
* PC card is inserted at the start of integration.
* PC card should not be removed during the integration.
* No communication is performed.
* Avoid excessive key operations.

Sufficient interval allows to save all data which corresponds to 1GB and 2GB capacity.
Top of this page
Adding such functions after unit purchasing is not available. Please be sure to add them at the time of main unit purchasing.
Top of this page
It is ±1V DC of the rated value for each range. Accuracy is ±(measurement accuracy + 0.2% f.s.), and output resistance is 22Ω±5%.
Top of this page

* Measurement cycle refers to the time interval used to measure
   the instantaneous real time value of one cycle of input voltage.

* Interval means the time interval for saving and output the measurement data.
* The r.m.s. value of voltage, current and electric power value in every one cycle
   are measured as an instantaneous value.
* Calculate the maximum, minimum and average values of instantaneous value being
   measured within the interval time and save them in the PC card in every interval time.
* Instantaneous value (for one cycle) of adjacent interval time is saved in the
   PC card in every interval time.

Top of this page
Header Characters in Appendix 2 File/Print Item Descriptions which is located in the instruction manual explains the details.
Top of this page
In case of three-phase three-wire three-current, the voltage of each phase electric power is phase voltage of virtual neutral point.
Voltage U1 is a line voltage so it is required to convert to phase voltage (line voltage /√3 ).
In this case, the phase voltage shall be 124.1V (215.0/√3).
Where, V1: 124.1V, I1:45.0A, P1:970W, the equation of power factor shall be 970/(124.1 x 45.0) = 0.174, hence, the measurement value is quite close.
Top of this page
Both basic measurement (voltage, current, electric power) and harmonic measurement are implemented based on PLL synchro system.Harmonic data is calculated based on the basic data ( voltage, current, electric power). Therefore, if the accuracy of basic data meets the specifications, then the harmonic measurement data is also approved to be within the specifications accordingly. This is the reason why no harmonic analysis item is presented in the calibration certificate.
Top of this page
A data save/ integration has been completed and the MIN, MAX, and AVE data must be cleared from the Measure List memory Go to the List screen from the Measure section. Push down the F5 button for 5 seconds, a "measurement data will be cleared proceed?" message will appear, press the enter button. You will now be able to change all configuration fields
Top of this page

Release of 2558A

2558A
2558A AC Voltage Current Standard New!

Details

Release of WT300 Series

WT300 Series
WT300 Series Digital Power Meters New!

Details

Release of DLM4000

DLM4000
DLM4000 series Mixed Signal Oscilloscope New!

Details

Release of CL420

CL420
Clamp-on Process Meter for DC mA Current #CL420 New!

Details

Release of WT1800

WT1800
WT1800 Precision Power Analyzer

Details

Release of DL850/DL850V

DL850/DL850V
DL850/DL850V Scopecorder

Details

Release of AQ6370C

AQ6370C
AQ6370C Optical Spectrum Analyzer

Details