FLXA21 Modular 2-Wire Dissolved Oxygen Analyzer

Models FLXA21
Modular 2-Wire Dissolved Oxygen Analyzer

Product Information

FLXA21 Modular 2-Wire Dissolved Oxygen Analyzer

The FLXA21 two-wire liquid analyzer offers single or dual sensor measurement. The modular-designed analyzer offers 4 kinds of measurements – pH/ORP (oxidation-reduction potential), contacting conductivity (SC), inductive conductivity (ISC) or dissolved oxygen (DO) – with the respective sensor module.

The FLXA21 Dissolved Oxygen Analyzer is a process analyzer with advanced functions and high reliability and requires less maintenance, so it is suitable for installation in the field and can even be used under harsh environments.
Not only a galvanic sensor but also a polarographic sensor can be used. It can display the dissolved oxygen concentration in units of mg/l, ppm, and percent saturation (%sat), and can make corrections for air pressure that changes with the altitude, etc., for the change in salt content of water, and for the change in temperature.

For the wide range of industrial environment, the FLXA21 Dissolved Oxygen Analyzer is designed with the enclosure of plastic, stainless steel or stainless steel with corrosion resistant coating.

   Bulletin for FLXA21
   General Specification for FLXA21
   General Specification for DO Sensor
   Technical Informarion for HART Communication 
   User's Manual for FLXA21

Featuers

* Dual sensor measurement on 2-wire type analyzer; DO sensor and DO sensor
* Calculated data from dual sensor measurement
* Redundant system on dual sensor measurement
* Easy touch screen operation on 2-wire type analyzer
* Simple HMI menu structure in 12 languages
(English, Chinese, Czech, French, German, Italian, Korean, Polish, Portuguese, Russian, Spanish and Japanese)
* Quick setup menu for immediate measurement
* On-line sensor checking and indication of sensor wellness
* HART communications
* Enclosure – plastic, stainless steel or stainless steel with corrosion-resistant coating

Specifications

Input specifications Input range is 0 to 50 μA for Galvanic sensors and 0 to 1 μA for polaro-graphic sensors. For temperature compensation the DO202 accepts Pt1000 (DO30G) and 22k NTC elements (OXYFERM and OXYGOLD sensors).
Input Range DO30 sensor:
  Dissolved Oxygen: 0 to 50 mg/l (ppm)
Oxyferm sensor:
  Measurement range: 10 ppb to 40 ppm
  Temperature range: 0 to 130℃
Oxygold G sensor:
  Measurement range: 2 ppb to 40 ppm
  Temperature range: 0 to 130℃
Oxygold B sensor:
  Measurement range: 8 ppb to 40 ppm
  Temperature range: 0 to 100℃
Output Range DO30 sensor:
  DO concentration :min.: 1 mg/l(ppm), maxi.: 50 mg/l(ppm)
  % saturation:min.: 10%, maxi.: 300%
  Temperature: -20 to 150℃
Oxyferm sensor:
  Measurement range: 10 ppb to 40 ppm
  Temperature range: 0 to 130℃
Oxygold G sensor:
  Measurement range: 2 ppb to 40 ppm
  Temperature range: 0 to 130℃
Oxygold B sensor:
  Measurement range: 8 ppb to 40 ppm
  Temperature range: 0 to 100℃
Temperature compensation Automatic using Pt1000 or 22k NTC or manual.
Range: 0 to 100℃ (32 to 212℉)
Calibration Semi-automatic one or two point calibration in air or in water using solubility tables of ISO 5814. The influence of pressure and salinity is taken into account when these specifications are entered. Also a two point manual calibration is possible.
Transmission Signals One output of 4 to 20 mA DC, Burn up (22 mA) or Burndwon (3.6 mA) to signal failure.
Communication mA-HART
Analyzer Structure Module structure
Housing Case:Plastic (Polycarbonate), stainless steel without painting, stainlesssteel with epoxy coating, stainless steel with urethane coating
Window:Polycarbonate (flexible)
Protection:NEMA4X, IP66
Environment and operation conditions Ambient Temperature: -20 to 55℃
Storage Temperature:-30 to +70℃
Humidity: 10 to 90% RH @ 40℃ (Non-condensing)
Regulatory Compliance Safety :EN61010-1, CSA C22.2 No.61010-1
EMC :EN61326-1 Class A,Table 2 (For use in industrial locations), EN61326-2-3, AS/NZS CISPR11, Korea Electromagnetic Conformity Standard
Installation altitude: 2000 m or less
Category based on IEC 61010: I
Pollution degree based on IEC 61010: 2
Explosion-proof ATEX, IECEx, FM, CSA, NEPSI
Power supply Nominal 24 V DC loop powered system
One (1) Sensor module (1 input): 16 to 40 V DC
Two (2) Sensor modules (2 inputs): 22.8 to 40 V DC
Conduit Adapter G1/2, 1/2NPT, M20 x 1.5
Weight Approx. 1 kg (Plastic housing), Approx. 2 kg (Stainless steel housing)

Performance (Accuracy)

(The specifications are expressed with simulated inputs.)

Performance in ppm mode Linearity: ±0.05 ppm or ±0.8% F.S., whichever is greater
Repeatability: ±0.05 ppm or ±0.8% F.S., whichever is greater
Accuracy: ±0.05 ppm or ±0.8% F.S., whichever is greater
Performance in ppb mode Linearity: ±1 ppb or ±0.8% F.S., whichever is greater
Repeatability: ±1 ppb or ±0.8% F.S., whichever is greater
Accuracy: ±1 ppb or ±0.8% F.S., whichever is greater
Temperature Linearity: ±0.3℃
Repeatability: ±0.1℃
Accuracy: ±0.3℃