Dual-Channel Conductivity Analyzer
모델코드 FLXA21 FLXA202 DC402G
지역 분류
  • General Purpose
  • Class I Div II (without the use of IS Barrier)
  • Class I Div I (with the use of IS Barrier)
  • General Purpose
  • Class I Div II (without the use of IS Barrier)*FM/CSA Pending
  • Class I Div I (with the use of IS Barrier)*FM/CSA Pending
  • General Purpose
  • Class I Div II
외함 재질 Plastic (Polycarbonate) Aluminum alloy die cast with epoxy coating (Standard offering)
Aluminum alloy die cast with urethane or high anti corrosion coating (available upon request)
Cast Aluminum with chemically resistance coating; polycarbonate cover with polycarbonate flexible window
하우징 등급 IP66/ NEMA 4A/ CSA 3S IP66 (except Canada), Type 4X (except Canada), Type 3S/4X (Canada) IP66/ NEMA 4A/ CSA 3S
전원 공급 2-wire 24VDC, Loop Powered 2-wire 24VDC, Loop Powered 4-Wire 110-256 VAC, Mains Power

출력

신호

Bi-directional HART digital communication, superimposed on mA (4-20mA) signal. Bi-directional HART digital communication, superimposed on mA (4-20mA) signal. Two isolated outputs of 4 to 20 mA DC with common negative.

Details

모델 FLEXA ™ 2 선식 분석기는 산업 설비의 연속 온라인 측정에 사용됩니다. 단일 또는 이중 센서 측정 옵션을 제공하므로 가장 유연한 2 선 분석기를 사용할 수 있습니다. 모델 FLEXA ™ 모듈 식 설계 시리즈 분석기입니다.

  • 하나의 분석기는 4 가지 측정 유형 중 하나를 수용 할 수 있습니다. pH / ORP, 접촉 전도도 (SC), 유도 전도도 (ISC) 및 용존 산소 (DO)
  • 2 선식 분석기 pH / ORP 및 pH / ORP, SC 및 SC, DO 및 DO에서 이중 센서 측정
  • 모듈 형 설계 : 교체 가능한 센서 모듈
  • 듀얼 센서 측정시 중복 시스템
  • 다양한 외함 재질 : SS 또는 플라스틱
  • 2 선식 분석기로 직관적인 간편한 터치 스크린 조작
  • 12 개 언어로 된 독창적 인 HMI 메뉴 구조
  • 빠르고 쉬운 측정 작업을위한 빠른 설정 메뉴
  • 예측 유지 보수를위한 온라인 센서 웰니스 검사
  • NEMA 4X / IP66 케이스
Input Specification

Two or four electrode measurement with square wave excitation, using max. 60 m cable and cell constants from 0.005 to 50.0 cm-1

Input Range Conductivity: 0 to 2000 mS/cm at 25ºC (77°F) 
Temperature: -20 to 140ºC (0 to 280°F)  
Conductivity 2 µS x K cm-1 to 200 mS x K cm-1
    
Accuracy:  ±0.5% F.S. 
 1 µS x K cm-1 to 2 µS x K cm-1     
      
Accuracy:  ±1% F.S. 
Resistivity 0.005kΩ/ K cm-1 to 0.5MΩ/ K cm-1
    
Accuracy:  ±0.5% F.S.
0.5MΩ/ x K cm-1 to 1MΩ/ x K cm-1
   
  Accuracy:  ±1% F.S. 
Temperature with Pt1000, Pb36, Ni 100
     Accuracy: ± 0.3ºC
with Pt100, NTC 8k55
      Accuracy: ± 0.4ºC
Step Response ≤7 seconds for 90% (for a 2 decade step)
Cable Length Max 60 meters (200 feet). 10 meters (35 feets0 fixed sensor cable + 50 meters (165 feet) WF10 extension cable
Transmission Signal Bi-directional HART digital communication, superimposed on mA (4-20 mA) signal
Temperature Compensation Automatic or manual, for temperature ranges.
Reference temp. programmable from 0-100°C or 30-210°F (default 25°C) 
Calibration Semi-automatic calibration using pre-configured OIML (KCl) standard tables, with automatic stability check. Manual adjustment to grab sample.
Display Black and white, 213 x 160 pixels LCD with a touch screen
Housing IP66/ NEMA 4A/ CSA 3S
IP66 (except Canada), Type 4X (except Canada), Type 3S/4X (Canada)
Environment Ambient Temperature: -20  to 55°C
Humidity: 10 to 90% RH at 40°C (non-condensing)
 Power Supply Ratings; 17 to 40 V DC 

Flexibility, reliability and low maintenance are among the benefits provided by the EXA DC402G conductivity analyzer. Designed to meet the exacting requirements of measuring dual cell conductivity and resistivity in the modern industrial environment, it contains many features to ensure the best precision whatever the application. Differential, ratio, deviation. % passage or % rejection can be indicated and/or transmitted. 

This 4-wire converter is housed in a robust IP65 field mountable case. Two mA outputs, four relays, digital communication and a clear LCD make the DC402G a truly comprehensive package. 

The DC402G features PI control on the auxilliary mA output and the pulse proportional relay outputs, thus avoiding the need for a separate controller.

The famous EXA sensor diagnostics are also present in the DC402G. Self-tuning of the excitation frequency and measuring pulses ensure optimum accuracy. The polarisation check gives on line indication of sensor fouling and early warning that maintenance is needed. A wide variety of temperature compensation possibilities (NaCl according to IEC 746-3, manual TC, preprogrammed matrices and a freely programmable 5x5 matrix), provides a high-accuracy measurement with minimum effort.

  • Differential, ratio or calculated outputs
  • Universal conductivity/resistivity, software switchable
  • On-line sensor checking
  • Self-tuning measuring signal
  • Matrix temperature compensation for pure water applications
  • Four fully configurable SPDT contact outputs
  • Two fully configurable mA outputs
  • Built-in PI controller
  • Easy to use EXA control panel
  • USP 23 monitoring

Downloads

동영상

기술 및 솔루션, 요꼬가와에 대한 자세한 정보를 찾고 계십니까?

견적 및 기술문의

Top