Limiting Current Type Oxygen Analyzer OX102

OX102 산소 분석기는 단일 분석기에서 ppm 에서 % 수준의 산소 농도를 측정하며 질소 리플 로우로 및 반도체 공장과 같은 응용 분야의 ppm 수준 측정에 이상적입니다.

지르코니아 산소 센서는 제한 전류 원리에 기반을두고 콤팩트하면서도 고성능을 유지하는 방법을 사용합니다. 센서는 직접 삽입으로 설치하도록 설계되었습니다.
샘플 가스가 질소 리플 로우로와 같은 응용 분야에서 수 ppm 수준의 유기 용매를 포함 할 경우 선택적 시료 샘플링 장치가 필요합니다.

Details

  • One instrument capable of measuring from 0 - 100 ppm to 0 - 25% oxygen
  • Simple one touch calibration
  • Self diagnostics assure high reliability
  • One touch sensor replacement
  • Compact size

Measuring Principle

Current flowing in the Zirconia fixed electrolyte causes oxygen to move from the negative to positive electrode (acts as an oxygen pump). If the flow of oxygen to the negative electrode is restricted by the alumina substrate then a current saturation area is created and the same current flows regardless of the applied voltage. The current that flows is called the limiting current. When a fixed voltage is applied, the current flow is proportional to the oxygen concentration.

 

 

Applications

  1. N2 reflow furnaces
  2. Semiconductor maker environmental monitoring
  3. ppm-level O2 measurement
  4. Monitoring purity of N2, Ar separated from air
  5. Monitoring for deficiency of O2
Measured gas O2 in N2 or mixed gases; cannot be used with silicone vapor, combustible gases and organic solvents
Measurement principle Zirconia limiting current method
Measuring range 0 - 100 ppm/0 - 1000 ppm/0 - 10000 ppm/0 - 25%, automatic/fixed switching mode
Sampling method Natural diffusion system. If the sampled gas may include organic solvents, use a sampling unit.
Sample gas conditions Temperature; 0 to 70℃
Humidity; 0 to 80%RH
Pressure; 1013±40 hPa(Abs)
Output signal 1. 4-20 mA DC isolated, fixed range(0 - 100, 0 - 1000, 0 - 1000 ppm, 0 - 25% selectable), or 4 - 20 mA DC autoranging
2. 1 to 5 V DC fixed range(0-100, 0-1000, 0-10000 ppm, 0-25% selectable)
3. Zone signal output for 4 - 20 mA DC autoranging
Contact output 2 outputs
Alarm output High or Low alarm
Calibration a. Atmospheric air one-point calibration : 15 to 25% O2 range
b. One-point calibration at 100 ppm : 80 to 120 ppm O2 range
c. One-point calibration at 1000 ppm : 900 to 1100 ppm O2 range
d. One-point calibration at 10000 ppm : 8500 to 9500 ppm O2 range
e. Complete calibration : with air, 20, 100, 500, 1000, 10000 ppm in that order
Extension cable 4 m, 9 m, 29 m
Power supply 100 to 240 V AC, 50/60 Hz, approx. 15 W
Converter size 96(W) x 96(H) x 190(D) mm
Repeatability ±15 ppm O2( Meas. range: 20 to 100ppm)
±25 ppm O2( Meas. range: 100 to 1000 ppm)
±0.3% of reading( Meas. range: 1000 to 5000 ppm)
±7.5% of reading( Meas. range: 5000 to 10000 ppm)
±0.25% O2( Meas. range: 1 to 25%)