UM33A-S SENCOM HMI Type

 
UM33A-S는 디지털 SENCOM 스마트 아답터 SA11에 연결된 센서와 사용 가능합니다. UM33A-S의 주요 기능은 센서의 측정 값을 표시하고 센서를 교정하는 것입니다. 

Details

Overview

간편함

  • 읽기 쉬운 14 세그먼트 대형 컬러 LCD 화면
  • 네비게이션 가이드 및 네비게이션 키
  • 스마트 아답터에 포함된 교정 기능 표시 및 제어
  • SENCOM용 전원공급 장치 : 3.6 - 5.5 VDC
  • 65-mm (depth)의 컴팩트한 디자인

신뢰성

  • 3년 보증*1
  • RoHS/WEEE
  • NEMA4*2/IP66 전면 패널

*1: 3 년 보증기간은 36개월입니다.
*2: Hose-down test only

사양

아이템 UM33A-S
입력 Input  Type pH, 전도도
PV 데이터 디스플레이   14-segment, 5 digits LCD display
디지털 출력 (option) Number
Function PV High, PV Low, etc
재전송 Number 1
Type pH, Conductivity
기타 사양 Power supply 100 to 240 VAC (+10%, -15%) or 24 VAC/DC (+10%, -15%) (option)
Power consumption Max 15 VA
Mass Max 0.5 kg

자료실

Standard the UM33A-S has a 100-240VAC power supply. Optional it is possible to order the 24VAC/DC power supply. ...
How to install the UM33A-S?
(an-um33a-s-03)
The UM33A-S is panel mounted. Tighten the screws with appropriate tightening torque within 0.25 Nm, otherwise it may cause the case deformation or the bracket will be damaged. Make sure that foreign materials do not enter the inside of the instrument...
What is the option /CV for?
(an-um33a-s-04)
This is to prevent anybody touching the power terminals (100-240VAC or 24VAC/DC).
Electro Magnetic Compatibility (CE, KC, ACMA): 61326-1 Class A, Table 2, EN 61326-2-3 EN 55011 Class A, Group 1 EN 61000-3-2 Class A EN 61000-3-3 Safety: IEC/EN 61010-1 (CE), IEC/EN 61010-2-201 (CE), IEC/EN 61010-2-030 (CE); CAN/CSA C22....
In case of a warning, the Event indicator will be lit, and by pressing the DISP key the Warning Code and scrolling message will be indicated. In case of more than 1 warning, the belonging Event indicators will be lit but only the Warning Code belongi...
Yes, it does. HOLD Display Active when HOLD function is set ON (active) in case of Calibration and/or Manual HOLD. LOCK lights up to indicate HOLD mode. HOLD during calibration When calibration is active the mA output can be set in a HOLD last (...
The display update cycle time can be changed to increase readability at lower temperatures.
It is possible to set a password in the UM33A-S. This password also can be changed when needed. When the password is forgotten it is no longer possible to access the device. Also, once activated it is not possible to de-activate the password.
Please see the drawing below of the terminal arrangement of the UM33A-S. To guarantee EMC- and Common Mode noise behavior of SA11, the FG terminal (#306) must be interconnected to Power supply Earth terminal (#112) by separate wire. Use WU11-S type c...
The main functions of the UM33A-S are: Displaying calculated main parameter values and WARN/ERROR statuses from SA11 Trigger calibration process of the sensor Generating mA process signal Built in power supply for the SA11 IP66/NEMA4X at the f...
The calibration limits are set in the SA11 and will be used during calibration and will generate an error when applicable. Limits can be set by using FieldMate or a FLXA402.
Default pre-programmed, see Appendix in Instruction Manual Configurable by expert HOST (= Yokogawa device FLXA402, Fieldmate, NOT UM33A-S)  Remark: some error configuration depends on the type of sensor that is connected. This will be autom...
The PH21 does not have a liquid Earth, hence does not measure ORP. When using a FU20 or FU24 ORP/REDOX will be measured, but will not show up in the display. Error messages can be configured for ORP and will give warnings. ...
The WU11 is to connect the SA11 to the FLXA402, UM33A-S, SMARTDAC+ or any other HOST system. WU11-…-V: for FU20F, FU24F, SC25F, and IB100 WU11-…-S: for SA11, BA11, IB100, and –VS sensors