DP320-8x
Mededeling: Dit product is niet meer leverbaar vanaf Dec 31, 2005.

Model Name: DP320-8x

Middelen

The exclamation point is displayed when data cannot be written to the hard disk. Check whether the save destination folder exists, and whether there is sufficient free space on the disk.
If you have not changed any settings, check the following. Check the amount of free space on the hard disk, and perform error checking Check the DA100 power, and whether the green STATUS lamp is lit Communication cable connection and replacement...
Yes, with the auto processor.
Since the OS security level has increased, you can no longer set the Ethernet module's IP address on Windows 7. Data acquisition will be unaffected, but for the IP address setting use a Windows XP PC with DAQ32 installed.
No. DAQ32Plus only supports 1 unit. We recommend using DAQLOGGER.
It is a number starting with 101, and is included in your DAQ32Plus CD case. If you lose the license number, order model no. WX102.
Upload settings as follows: Darwin setting icon > Communication > Receive. Next, select Communication > Execute > Balance. Select Balance for the execute type, then select the range of channels and execute the balancing. (Setting the exec...
Use the latest version of SMARTDAC+ Standard Universal Viewer.
Ethernet, RS-232C, RS-422A/485, and GPIB.
Communication cannot open if the IP address is the same.
Launch the DAQ32Plus monitor, then start the client.
Yes, by using the DAQ32Plus client package. The latest version supports monitoring of 1 DAQ32Plus monitor server by up to 16 clients.
Up to 30 groups of up to 32 channels each can be created. You can set the display span in the detailed display settings.
Check whether "Alarm data" is checked in the DAQ32Plus Logger acquisition conditions. If not, the alarm on the DAQ32Plus screen remains green and doesn't turn red even if an alarm is reported on the DARWIN hardware.
It can monitor up to 20 channels. If you need more channels, please consider DAQ32Plus (supports up to 300 channels).
The time in the trends is the time on the DA100. When connected for a long time, the times on the PC and DA100 can start to diverge. Closing and then reopening the connection will synchronize the times because the PC's time is automatically set o...
Double-click the DAQ32Plus environment settings icon, click the Environment tab, and make sure ETHERNET is selected and that the DA100's IP address is set. If the IP address is not set, set it and then restart DAQ32Plus.
Because DAQ32 Plus has greater functionality than DAQ32, it requires that much more CPU speed and memory. If there was no problem on DAQ32 but the performance bar goes to OVER on DAQ32 Plus, probably the baud rate set on DAQ32 is too slow for DAQ32 P...
You can, starting with style 8 of the DA. Setting command: VS. Control execute command: VD. The DR and DC do not support this. With DAQ32 and DAQ32Plus you can set up output relays, but you can't actually operate them with commands. You can also ...
On the RS232C communication module, set communication parameters by using the dip switches on the front of the module. Make sure the communication module’s dip switches match what is displayed on the DAQ32's “Environment, diagnosis, a...
For 0.5 second models, temperature is measured for RJC every 2.5 seconds (with a 0.5 second scan, since the time for processing temperature measurement, burnout, and other factors is too short, measurement occurs every 5 cycles). For 2.0 second mode...
Generally they are not initialized if the modules are not different from the previous system. However, initialization happens if slots are in the following conditions. Slot replaced with a module of a different model name or style number Slot who...
For voltage measurement, if 5% of range + range span is exceeded, an overrange occurs (±32767 is output). For temperature range, the +10°C measuring range is the upper limit. Ex) For the ±50 V range (span of -50 to +50), the upper l...
Does DARWIN support Windows XP?
(ns-faq-darwin-8017-spec)
R10.03 and later support Windows XP.
The voltage generated when a burnout current of 4.2 mA flows to the measuring terminal circuit is measured in the 60 mV range, and a determination is made. If 42 mV (10 kΩ) or more, the result is 0 (open). If less than that, 1 (closed).
DAQ32 (DP120) or DAQWORX DAQ32 Plus (WX102) installed on Windows 7 or Windows Vista cannot change the IP address of DARWIN series Ethernet modules (suffix code /C7) or DARWIN series standalone models with the Ethernet communications option (model cod...
Yes. Set the IP address, subnet mask, and default gateway on the main unit.
The Viewer cannot determine the numbers after the decimal with statistical computation on DI input. Data acquired and results computed by the BFD32 are affected by the decimal of the specified measurement range. For DI (contact input), a 4.2 µA...