UM33A-S SENCOM HMI Type

 
The UM33A-S connects to sensors which are connected to a digital SENCOM Smart Adapter, model SA11. The main functions of the UM33A-S are displaying the measured values of the sensor and calibrating the sensor.

Details

Overview

SimplicitySENCOM system: UM33A-S + SA11

  • Easy-to-read, 14-segment large color LCD display
  • Navigation guide & navigation keys
  • Indication and control of the calibration function included in the Smart Adapter
  • Power supply for SENCOM: 3.6 - 5.5 VDC
  • Compact design of 65-mm depth

Reliability

  • 3 year warranty*1
  • RoHS/WEEE
  • NEMA4*2/IP66 front panel

*1: The 3 year warranty extendes 36 months
*2: Hose-down test only

Specifications

ITEM UM33A-S
Input Input type pH, Conductivity
PV data display   14-segment, 5 digits LCD display
Digital outputs (option) Number
Function PV High, PV Low, etc
Retransmission Number 1
Type pH, Conductivity
Other specifications Power supply 100 to 240 VAC (+10%, -15%) or 24 VAC/DC (+10%, -15%) (option)
Power consumption Max 15 VA
Mass Max 0.5 kg

Resources

Standard the UM33A-S has a 100-240VAC power supply. Optional it is possible to order the 24VAC/DC power supply. ...
How to install the UM33A-S?
(an-um33a-s-03)
The UM33A-S is panel mounted. Tighten the screws with appropriate tightening torque within 0.25 Nm, otherwise it may cause the case deformation or the bracket will be damaged. Make sure that foreign materials do not enter the inside of the instrument...
What is the option /CV for?
(an-um33a-s-04)
This is to prevent anybody touching the power terminals (100-240VAC or 24VAC/DC).
Electro Magnetic Compatibility (CE, KC, ACMA): 61326-1 Class A, Table 2, EN 61326-2-3 EN 55011 Class A, Group 1 EN 61000-3-2 Class A EN 61000-3-3 Safety: IEC/EN 61010-1 (CE), IEC/EN 61010-2-201 (CE), IEC/EN 61010-2-030 (CE); CAN/CSA C22....
In case of a warning, the Event indicator will be lit, and by pressing the DISP key the Warning Code and scrolling message will be indicated. In case of more than 1 warning, the belonging Event indicators will be lit but only the Warning Code belongi...
Yes, it does. HOLD Display Active when HOLD function is set ON (active) in case of Calibration and/or Manual HOLD. LOCK lights up to indicate HOLD mode. HOLD during calibration When calibration is active the mA output can be set in a HOLD last (...
The display update cycle time can be changed to increase readability at lower temperatures.
It is possible to set a password in the UM33A-S. This password also can be changed when needed. When the password is forgotten it is no longer possible to access the device. Also, once activated it is not possible to de-activate the password.
Please see the drawing below of the terminal arrangement of the UM33A-S. To guarantee EMC- and Common Mode noise behavior of SA11, the FG terminal (#306) must be interconnected to Power supply Earth terminal (#112) by separate wire. Use WU11-S type c...
The main functions of the UM33A-S are: Displaying calculated main parameter values and WARN/ERROR statuses from SA11 Trigger calibration process of the sensor Generating mA process signal Built in power supply for the SA11 IP66/NEMA4X at the f...
The calibration limits are set in the SA11 and will be used during calibration and will generate an error when applicable. Limits can be set by using FieldMate or a FLXA402.
Default pre-programmed, see Appendix in Instruction Manual Configurable by expert HOST (= Yokogawa device FLXA402, Fieldmate, NOT UM33A-S)  Remark: some error configuration depends on the type of sensor that is connected. This will be autom...
The PH21 does not have a liquid Earth, hence does not measure ORP. When using a FU20 or FU24 ORP/REDOX will be measured, but will not show up in the display. Error messages can be configured for ORP and will give warnings. ...
The WU11 is to connect the SA11 to the FLXA402, UM33A-S, SMARTDAC+ or any other HOST system. WU11-…-V: for FU20F, FU24F, SC25F, and IB100 WU11-…-S: for SA11, BA11, IB100, and –VS sensors

Downloads

Looking for more information on our people, technology and solutions?

Contact Us

Top