Electrolysis Plant Trace H2O (Moisture) Measurement by the TDLS200

下載 (993 KB)

Introduction

If water is present after the dry chlorine tower in brine electrolysis plants, the downstream compressor is corroded or the product quality is deteriorated. The moisture level is measured before and after the compressor to prevent the corrosion of the compressor and the deterioration of quality. Conventional trace moisture analyzers using a phosphorus pentoxide (P2O5) sensor measure process gas with the contacting sensor so there are problems such as the deterioration of the sensor, slow response time, high maintenance, and high running costs. The TDLS200 Laser Analyzer is the solution to these problems.

Process Overview

Chlorine gas generated in the brine electrolysis plant is dehumidified in the wet chlorine tower and sent to the drying tower where the chlorine gas is dried using sulfuric acid solution. The dried chloride gas is sent to the dry chloride tower where the gas is dehumidified, and finally, the dry chloride gas is compressed by the compressor and liquefied, and then shipped as a product.

Expected Benefits

  • Reliable measurement with less deterioration by non-contact measurement
  • Ability to measure even corrosive gasses with less deterioration reduces the maintenance and running costs
  • Fast response to and stable measurement of changes in H2O concentration

Electrolysis Plant - Exterior View of TDLS200

Electrolysis Plant (Exterior View of TDLS200 Trace Moisture Analyzer)

Solution Details

Unlike analyzers using a phosphorus pentoxide (P2O5) sensor, the TDLS responds fast to changes in the trace moisture of the sample.

Field Data

Configuration Example of TDLS200 Laser Analyzer

Responsiveness Comparison between the TDLS200 and Moisture Analyzer Using P2O5 Sensor

Configuration Example of TDLS200 Laser Analyzer Trace Moisture Measurement

A sampling system is used to introduce a process gas into the flow cell to measure it.

Electrolysis Plant Trace H2O Measurement

Features

  • A proprietary sampling method enables a high-sensitivity measurement (in the minimum range of 0-30 ppm)
  • In-situ calibration is possible without removing the system
  • The sampling system minimizes the effect ofmoisture in the purge gas

TDLS Analyzer

  • TDLS200
    Available on a special order basis.

Notes

  • Power supply: 100 - 240 V AC, 50/60 Hz
  • Purge gas: Instrumentation and N2
    Inside the sampling board and the light emitter and light receiver boxes
    Purge: Instrumentation air 10 L/min
    Laser light transmission region purge: N2 flow rate 5 L/min

產業別

相關產品&解決方案

In-Situ Gas Analyzer TDLS200

The TDLS200 is designed to make fast, accurate measurements of gases including oxygen (O2), carbon monoxide (CO), methane (CH4), moisture (H2O), carbon dioxide (CO2), ammonia (NH3), and hydrogen chloride (HCl). It is ideally suited for environments involving changing pressure or temperature.

TDLS8000在線氣體分析儀

橫河電機的新TDLS™8000功能強大,將諸多先進工業特性集成於一台耐用設備中。其平台設計適用於現場測量,無需進行樣品抽取和樣品處理。

氧氣濃度計

氧氣濃度計為燃燒控制、製程質量、安全和環境應用提供了有價值的測量數據。

×

Have Questions?

Contact a Yokogawa Expert to learn how we can help you solve your challenges.

置頂