Introduction:
Wastes have been considered to be a serious worldwide environmental problem in recent years. Because of increasing pollution, these wastes should be treated. However, industrial wastes can contain a number of valuable organic components. Recovery of these components is important economically. Using conventional distillation techniques, the separation of acetic acid and water is both impractical and uneconomical, because it often requires large number of trays and a high reflux ratio. In practice special techniques are used depending on the concentration of acetic acid.
Between 30 and 70% (w/w) acetic acid contents, extractive distillation was suggested. Extractive distillation is a multicomponent-rectification method similar in purpose to azeotropic distillation. In extractive distillation, to a binary mixture which is difficult or impossible to separate by ordinary means, a third component termed an entrainer is added which alters the relative volatility of the original constituents, thus permitting the separation. In our department acetic acid is used as a solvent during the obtaining of cobalt(III) acetate from cobalt(II) acetate by an electrochemical method.
After the operation, the remaining waste contains acetic acid. In this work, acetic acid which has been found in this waste was recovered by extractive distillation. Adiponitrile and sulfolane were used as high boiling solvents and the effects of solvent feed rate/solution feed rate ratio and type were investigated. According to the experimental results, it was seem that the recovery of acetic acid from waste streams is possible by extractive distillation.
There are many methods available for the separation of acetic acid like, solvent extraction, distillation, adsorption, pervaporation and ion exchange.
Why pH is important?
If we take one of the separation technique as an example say using liquid membrane for separation. The vegetable oil being hydrophobic in nature allows the selective transport of acetic acid through it from the feed phase to the strip phase. Acetic acid is transported across the membrane from the feed phase in its unionized molecular form and not in its ionic form that is acetate. This required that the pH of the feed phase is lower than the pKa value of acetic acid which is 4.76 (at 25 degC). At pH above 4.76 acetate ion is formed, which cannot be permeate through the liquid membrane.
The pH value of the feed phases of 0.1 M, 0.05 M and 0.01 M concentrations of acetic acid was found to be 3.23, 3.65 and 4.05 respectively. These pH values are lower than the pKa value of acetic acid, enabling permeation of acetic acid across the membrane.
Critical area: Pure acetic acid service, which will have low conductivity. Acetic acid (systematically named ethanoic acid is an organic compound with the chemical formula CH3COOH (also written as CH3CO2H or C2H4O2). It is a colorless liquid that when undiluted is also called glacial acetic acid. Acetic acid has a distinctive sour taste and pungent smell. Although it is classified as a weak acid, concentrated acetic acid is corrosive. Acetic acid is the second simplest carboxylic acid (the simplest is formic acid) and is an important chemical reagent and industrial chemical.
Typical process details:
Pure acetic acid service, more than 30%
Water even less than <1%.
Typical problems:
The acid has very low conductivity so measurement is difficult.
Remedies:
To use a flowing special sensor to get a good electrolytical contact between reference cell and process.
Product Recommendations
Measurement System
Process Liquid Analyzer:
- 2-Wire Analyzer FLXA202
Dual sensor measurement on 2-wire type analyser
Redundant system on dual sensor measurement
Easy touch screen operation on 2-wire type analyzer
- 4-Wire Converter FLXA402
Connectable to upto five sensor
Easily viawable color LCD
Touch screen operation
Sensor Selection:
Yokogawa sensor SC21C-AGC55 can be the solution.
This flowing electrolyte sensors uses a ceramic reference junction, with the electrolyte (3.3 molal KCl). The flow of electrolyte through the junction, while small, remains the safest way to prevent clogging and to protect the internal reference against poisoning and diffusion.
Features for type SC21C-AGC55
- For tough application where pollution of the reference system is to be expected
- Low ionic application where the positive flow of electrolyte provides the conductivity needed to measure pH(< 50 µS/cm)
- Heavy duty pH sensitive glass.
- Flowing reference system for pollution resistance, and highly stable reference potential.
- Use in combination with the presurisable electrolyte reservoir to obtain a positive flow towards the process (K1500YA)
Temperature Sensor
SM60-T1 SERIES - For accurate pH measurement temperature compensation is required. Either a Pt100 or a Pt 1000 temperature electrode can be selected.
Cables used:
WU20-LT** Qty1 + WU20-PC** Qty1
Flow fitting:
F*20 various type of flow fittings can be used
Tangible benefit
Save down time in cleaning, repeated calibration, improve end product quality.
Note: For additional information on this application contact the local Yokogawa Process Liquid Analyzer Department
Industrien
-
Petrochemical
Petrochemical companies’ needs are extremely diverse. To come out ahead in today’s highly competitive marketplace, producers are striving to improve quality and productivity. Yokogawa provides tailor-made solutions for these needs based on its long and wide-ranging experience in this field.
Zugehörige Produkte & Lösungen
-
2-Leiter-Analysator FLXA202/21
Die Analysatoren der FLEXA-Serie werden für kontinuierliche Online-Messungen in industriellen Anlagen verwendet. Mit der Option für die Ein- oder Zwei-Sensor-Messung sind sie die flexibelsten 2-Leiter-Analysegeräte auf dem Markt.
-
2-Leiter-Messumformer/-Analysator FLXA202
Die Analysatoren der FLEXA™ Serie werden für kontinuierliche Online-Messungen in industriellen Anlagen verwendet. Mit der Option für die Ein- oder Zwei-Sensor-Messung sind sie die flexibelsten 2-Leiter-Analysegeräte auf dem Markt.
-
Digitaler SENCOM™ SMART-Adapter, SA11
Wiederverwendbarer SMART-Adapter, bei dem nur der Analogsensor am Ende der Lebensdauer entsorgt werden muss. Mit der Plattform SENCOM 4.0 reduziert Yokogawa Kosten und Abfälle und trägt damit zu seinem langfristigen Unternehmensziel einer nachhaltigen Zukunft für alle bei.
-
Durchfluss/NPT-Armaturen FF20/FS20
Yokogawa hat viel Aufwand und Zeit in die Entwicklung einer breiten Palette von Armaturen investiert. Ein besonderer Schwerpunkt lag dabei auf der Senkung von Installations- und Wartungskosten und somit günstigeren Betriebskosten.
-
Industrieelektroden für pH/Redox-Messungen
Das Herzstück einer pH-Messschleife ist das Elektrodensystem. Yokogawa hat ein umfangreiches Sortiment von Elektroden entwickelt, die dafür sorgen, dass dieses Herz auch unter den schwierigsten Bedingungen weiter schlägt.
-
Mehrkanal-4-Leiter-Analysator FLXA402
Die Analysatoren der FLEXA™-Serie sind modular gestaltete Analysegeräte für kontinuierliche Online-Messungen in industriellen Anlagen. Sie ermöglichen Messungen mit einem oder mehreren Sensoren.
-
pH- und ORP-Analysatoren
Sensoren und Analysatoren für pH und ORP (Redox) werden zur kontinuierlichen Überwachung von pH- und ORP-Werten im Prozess eingesetzt, z.B. um die Wasser- und Produktqualität sicherzustellen oder um Abflüsse, Batch-Neutralisierung, Zellstoffbestand, Gaswäscher, Kühltürme, chemische Aufbereitung, Wasser- und Abwasserbehandlung und viele andere Anwendungen zu überwachen.
-
pH- und Redoxsensoren
pH-Elektroden und -Sensoren sind die detektierenden Elemente bei der Messung des pH-Werts. Die verschiedenen Installationsoptionen schließen u. a. Wechsel-, Durchfluss- und Eintaucharmaturen sowie den direkten Einbau ein. Die richtige Auswahl von pH-Elektroden und Sensoren ist für optimale Messergebnisse von entscheidender Bedeutung.