Source measure cycle (source & measure at 1 ms minimum)
As shown in the figure below, the GS610 is capable of synchronizing
the generation and measurement at any operation mode. Generation
is started after a source delay with respect to an internal/external
trigger. Then, after the time it takes for the source level to stabilize
(measure delay) the measurement is carried out over a given integration
time. This function is necessary when making measurements by waiting
for the load response after applying a pulse. The source measure
cycle is the minimum unit for signal generation and measurement.
The source/measurement operation (source measurement cycle) on the
GS610 is 1 ms minimum.
Difference measurement using the sweep operation
The sweep operation can be used to make two measurements, then
the difference between the two measurements can be determined. When
the source value of sweep data 1 is established, the first measurement
is carried out. Then, when the source value of sweep data 2 is established,
the second measurement is carried out. The difference is determined
by subtracting the value of the second measurement from that of the
first measurement. By carrying out the two measurements within a short
time, highly accurate measurements are possible with the effects of
offset fluctuation cancelled. In addition, by applying a voltage
with equal magnitude but opposite in polarity (source 1 = -1×
source 2), the effects of offset error caused by the thermo electromotive
force at the junction can be cancelled.