N Single Mode
When you need to observe closely-spaced waveform events consecutively,
it is important to minimize the dead time between captures. The
'N single' mode on the DL9000 captures up to 1,600 waveforms on each
of 4 channels with as little as 400 ns of dead time between acquisitions.
In the 500 ps/div range, this corresponds to an effective acquisition
rate of 2.5 M waveforms/sec/ch.
New ACCUM (accumulation) mode
When observing long-term repetitive waveform events, the ACCUM function
offers a fast repetitive signal acquisition rate of up to 25 k waveforms/sec/ch
(on 4 channels simultaneously) while retaining up to 2,000 acquisitions
in memory.
Whether you use N single or the ACCUM function, previously acquired
waveforms are stored in memory and can be accessed using the History
Function.