Product Tutorials
Power Meter Tutorials
Calibration
Pwer Meter Calibration
Traceability of wide-bandwidth power meter
Calibration method for the working standard
- Voltage calibration
- Current calibration
- Phase angle measurement of the capacitor
- Preparing some capacitors each frequency (50/ 60/ 400/ 1k/ 2k/ 5k / 10k/ 20k/ 50k/ 100k/ 200k/ 500kHz)
- Measuring admittance (Conductance-G and susceptance-X) of the capacitor using precision LCR meter
- Calculating the phase angle of the capacitor
- Power calibration at PF=0
(It measures Internal phase difference between voltage and current input circuit.)
- Power calibration at PF=1
Calibration Method for working standard at PF=0
Capacitor's Equivalent Circuit
Consumption power of the resistor in the capacitor
Working standard displays power value (expectation)
Compensate of the input impedance of the voltage input circuit
Phase difference (_t) including the influence of input impedance of the voltage input circuit
Consumption power of the total admittance
Working standard displays power value