If input RMS value is lower than rated range value, the power meter can measure signal with more large crest factor. Below table shows between available input signal and the crest factor.

Input signal range which is defined accuracy, is within 10 to 110% of rated range. Therefore, in CF=3, the crest factor as the size of dynamic range for the WT series is up to 30 (at 10% of rated range value input) within the range of defined accuracy.
For your information, the WT series can display value when it puts the signal that is more than 0.5% of rated range value. If the input RMS value is 0.5% of rated range and input peek value is three times larger than rated range value, the crest factor becomes 600.
If an input exceeding the value shown in the specifications is measured, the peak area of the waveform is clipped to cause an error.