| Object of measurement |
Conductivity of liquids or solutions |
| Measuring Principle |
2-electrode or 4-electrode system |
| Cell Constants |
0.01 cm-1 or 10 cm-1 (for 2-electrode system)
10 cm-1 (for 4-electrode system) |
| Measuring Ranges |
0 - 0.5 μS/cm to 0 - 100 μS/cm for a cell constant of 0.01 cm-1
0 - 1 mS/cm to 0 - 1000 mS/cm for a cell constant of 10 cm-1 |
| Temperature Range |
0 to 100 deg C (130 deg C maximum only for 0.01 cm-1 cell constant detectors, excluding those with polypropylene chambers) |
| Pressure |
1000 KPa max. (500 kPa maximum for detectors with polypropylene chambers) |
| Flow rate of Sample Solution |
No particular limitation applies, although a value of less than 20 l/min is recommended for flow-through detectors. |
| RTD for Temperature Compensation |
Pt1000 (built into the sensor) |
| Construction |
Rainproof encapsulation (compatible with JIS C0920) |
| Weight |
Screw-in: approx. 1.3 kg (without the cable)
Flow-through (SCS14 chamber): approx. 3.1 kg (without the cable)
Flow-through (SCS14 chamber, flanged): approx. 4.5 kg (without the cable)
Flow-through (polypropylene chamber): approx. 2.7 kg (without the cable)
Flow-through (polypropylene chamber, flanged): approx. 3.2 kg (without the cable)
Cable: 0.3 kg for 5.5 m length; 0.5 kg for 10 m length |
| Process Connection |
Screw-in or flow-through |